Your browser doesn't support javascript.
loading
The influence of patient positioning and immobilization equipment on MR image quality and image registration in radiation therapy.
Rostami, Atefeh; Robatjazi, Mostafa; Javadinia, Seyed Alireza; Shomoossi, Nematullah; Shahraini, Ramin.
Afiliación
  • Rostami A; Department of Medical Physics and Radiological Sciences, Sabzevar University of Medical Sciences, Sabzevar, Iran.
  • Robatjazi M; Department of Medical Physics and Radiological Sciences, Sabzevar University of Medical Sciences, Sabzevar, Iran.
  • Javadinia SA; Non-Communicable Diseases Research Center, Sabzevar University of Medical Sciences, Sabzevar, Iran.
  • Shomoossi N; Non-Communicable Diseases Research Center, Sabzevar University of Medical Sciences, Sabzevar, Iran.
  • Shahraini R; School of Medicine, Sabzevar University of Medical Sciences, Sabzevar, Iran.
J Appl Clin Med Phys ; 25(2): e14162, 2024 Feb.
Article en En | MEDLINE | ID: mdl-37716368
INTRODUCTION: MRI is preferred for brain tumor assessment, while CT is used for radiotherapy simulation. This study evaluated immobilization equipment's impact on CT-MRI registration accuracy and MR image quality in RT setup. METHODS: We included CT and MR images from 11 patients with high-grade glioma, all of whom were immobilized with a thermoplastic mask and headrest. T1- and T2-weighted MR images were acquired using an MR head coil in a diagnostic setup (DS) and a body matrix coil in RT setup. To assess MR image quality, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were considered in some dedicated regions of interest. We also evaluated the impact of immobilization equipment on CT-MRI rigid registration using line profile and external contour methods. RESULTS: The CNR and SNR reduction was in the RT setup of imaging. This was more evident in T1-weighted images than in T2-weighted ones. The SNR decreased by 14.91% and 12.09%, while CNR decreased by 25.12% and 20.15% in T1- and T2-weighted images, respectively. The immobilization equipment in the RT setup decreased the mean error in rigid registration by 1.02 mm. The external contour method yielded Dice similarity coefficients (DSC) of 0.84 and 0.92 for CT-DS MRI and CT-RT MRI registration, respectively. CONCLUSION: The image quality reduction in the RT setup was due to the imaged region's anatomy and its position relative to the applied coil. Furthermore, optimizing the pulse sequence is crucial for MR imaging in RT applications. Although the use of immobilization equipment may decrease the image quality in the RT setup, it does not affect organ delineation, and the image quality is still satisfactory for this purpose. Also, the use of immobilization equipment in the RT setup has increased registration accuracy.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Radioterapia Guiada por Imagen Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article País de afiliación: Irán Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Radioterapia Guiada por Imagen Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article País de afiliación: Irán Pais de publicación: Estados Unidos