Subject-specific timing adaption in time-encoded arterial spin labeling imaging.
Breutigam, Nora-Josefin; Hoinkiss, Daniel Christopher; Konstandin, Simon; Buck, Mareike Alicja; Mahroo, Amnah; Eickel, Klaus; von Samson-Himmelstjerna, Federico; Günther, Matthias.
Afiliación
  • Breutigam NJ; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany. nora-josefin.breutigam@mevis.fraunhofer.de.
  • Hoinkiss DC; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany.
  • Konstandin S; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany.
  • Buck MA; Mediri GmbH, Heidelberg, Germany.
  • Mahroo A; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany.
  • Eickel K; Faculty 1 (Physics/Electrical Engineering), University of Bremen, Bremen, Germany.
  • von Samson-Himmelstjerna F; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany.
  • Günther M; Imaging Physics, Fraunhofer Institute for Digital Medicine MEVIS, Max-von-Laue-Str. 2, 28359, Bremen, Germany.
MAGMA ; 37(1): 53-68, 2024 Feb.
Article en En | MEDLINE | ID: mdl-37768433
ABSTRACT

OBJECTIVES:

One challenge in arterial spin labeling (ASL) is the high variability of arterial transit times (ATT), which causes associated arterial transit delay (ATD) artifacts. In patients with pathological changes, these artifacts occur when post-labeling delay (PLD) and bolus durations are not optimally matched to the subject, resulting in difficult quantification of cerebral blood flow (CBF) and ATT. This is also true for the free lunch approach in Hadamard-encoded pseudocontinuous ASL (H-pCASL). MATERIAL AND

METHODS:

Five healthy volunteers were scanned with a 3 T MR-system. pCASL-subbolus timing was adjusted individually by the developed adaptive Walsh-ordered pCASL sequence and an automatic feedback algorithm. The quantification results for CBF and ATT and the respective standard deviations were compared with results obtained using recommended timings and intentionally suboptimal timings.

RESULTS:

The algorithm individually adjusted the pCASL-subbolus PLD for each subject within the range of recommended timing for healthy subjects, with a mean intra-subject adjustment deviation of 47.15 ms for single-shot and 44.5 ms for segmented acquisition in three repetitions.

DISCUSSION:

A first positive assessment of the results was performed on healthy volunteers. The extent to which the results can be transferred to patients and are of benefit must be investigated in follow-up studies.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Arterias / Imagen por Resonancia Magnética Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: MAGMA Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Arterias / Imagen por Resonancia Magnética Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: MAGMA Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article País de afiliación: Alemania