Motion Characterization of Pacemaker Lead Wire In Vivo for Piezoelectric Energy Harvesting Applications.
Hu, Christopher; Behdinan, Kamran.
Afiliación
  • Hu C; Advanced Research Laboratory for Multifunctional Lightweight Structures (ARL-MLS), Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, ON, M5S 3G8, Canada.
  • Behdinan K; Advanced Research Laboratory for Multifunctional Lightweight Structures (ARL-MLS), Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, ON, M5S 3G8, Canada. behdinan@mie.utoronto.ca.
Article en En | MEDLINE | ID: mdl-37991598
PURPOSE: Piezoelectric energy harvesters (PEH) for cardiac pacemakers typically use animal models to assess the performance of the PEH. However, if considering multiple designs, the use of animal models and prototyping increases costs and time. To reduce the use of animal models in research for pacemaker energy harvesting applications, this study investigates the motion of a pacemaker lead wire (PLW) in vivo using fluoroscopy imaging to quantify the position and displacements as a function of time, such that the data can be used in computer simulations. METHODS: The proposed technique uses fluoroscopy imaging video data of a dual chamber pacemaker implanted in a patient, and image processing allows for the motion of the PLW captured. The motion is discretized into nodes for ease of implementation in finite element software. FEA simulation is presented using a piezoelectric energy harvester design integrated in the lead wire, and the energy output is predicted by finite element computer simulation. RESULTS: A 2-dimensional analysis is conducted with the fluoroscopy imaging video data to characterize the PLW motion and results show close agreement with literature values. Simulations with an energy harvesting circuit using the nodal position and displacement data shows that a PEH integrated in the PLW can generate a direct current voltage of 1.12 V and power output of 0.125 µW, potentially extending the battery life of pacemakers by 0.75-1 years. CONCLUSIONS: The results suggest that fluoroscopy imaging data can be effective in evaluating PEH designs rather than using animal models, saving time and costs.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Cardiovasc Eng Technol Año: 2023 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Cardiovasc Eng Technol Año: 2023 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos