Your browser doesn't support javascript.
loading
Enhanced Precision in Genioplasty: A Novel Intraoperative Spatial Repositioning Using Computer-Aided Design and Manufacturing Technology and a Holographic Mixed Reality Application.
Koyachi, Masahide; Sugahara, Keisuke; Tachizawa, Kotaro; Nishiyama, Akihiro; Odaka, Kento; Matsunaga, Satoru; Sugimoto, Maki; Tachiki, Chie; Nishii, Yasushi; Katakura, Akira.
Afiliación
  • Koyachi M; Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Sugahara K; Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Tachizawa K; Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Nishiyama A; Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Odaka K; Department of Oral and Maxillofacial Radiology, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Matsunaga S; Department of Anatomy, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Sugimoto M; Department of Oral Pathobiological Science and Surgery, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Tachiki C; Innovation Lab, Teikyo University Okinaga Research Institute, 2-16-1 Hirakawacho Chiyoda-ku, Tokyo 102-0093, Japan.
  • Nishii Y; Department of Orthodontics, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
  • Katakura A; Department of Orthodontics, Tokyo Dental College, 2-9-18 Kandamisaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.
J Clin Med ; 12(23)2023 Nov 29.
Article en En | MEDLINE | ID: mdl-38068460
ABSTRACT
Genioplasty is performed for the orthognathic surgical correction of dentofacial deformities. This article reports a safe and accurate method for genioplasty combining a novel three-dimensional (3D) device with mixed reality (MR)-assisted surgery using a registration marker and a head-mounted display. Four types of devices were designed based on the virtual operation a surgical splint with a connector; an osteotomy device; a repositioning device; and a registration marker. Microsoft HoloLens 2 and Holoeyes MD were used to project holograms created using computed tomography (CT) data onto the surgical field to improve the accuracy of the computer-aided designed and manufactured (CAD/CAM) surgical guides. After making an incision on the oral vestibule, the splint was fitted on the teeth and the osteotomy device was mounted at the junction site, placed directly on the exposed mandible bone surface. Temporary screws were fixed into the screw hole. An ultrasonic cutting instrument was used for the osteotomy. After separating the bone, a repositioning device was connected to the splint junction and bone segment, and repositioning was performed. At the time of repositioning, the registration marker was connected to the splint junction, and mandible repositioning was confirmed three-dimensionally through HoloLens 2 into the position specified in the virtual surgery. The rate of overlay error between the preoperative virtual operation and one-month postoperative CT data within 2 mm was 100%. CAD/CAM combined with MR enabled accurate genioplasty.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Clin Med Año: 2023 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Clin Med Año: 2023 Tipo del documento: Article País de afiliación: Japón