Your browser doesn't support javascript.
loading
Pushing MP2RAGE boundaries: Ultimate time-efficient parameterization combined with exhaustive T1 synthetic contrasts.
Bapst, Blanche; Massire, Aurélien; Mauconduit, Franck; Gras, Vincent; Boulant, Nicolas; Dufour, Juliette; Bodini, Benedetta; Stankoff, Bruno; Luciani, Alain; Vignaud, Alexandre.
Afiliación
  • Bapst B; University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France.
  • Massire A; Department of Neuroradiology, AP-HP, Henri Mondor University Hospital, Créteil, France.
  • Mauconduit F; EA 4391, Université Paris Est Créteil, Créteil, France.
  • Gras V; Siemens Healthcare SAS, Courbevoie, France.
  • Boulant N; University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France.
  • Dufour J; University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France.
  • Bodini B; University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France.
  • Stankoff B; Sorbonne Université, Paris Brain Institute, ICM, CNRS, Inserm, Paris, France.
  • Luciani A; Sorbonne Université, Paris Brain Institute, ICM, CNRS, Inserm, Paris, France.
  • Vignaud A; Sorbonne Université, Paris Brain Institute, ICM, CNRS, Inserm, Paris, France.
Magn Reson Med ; 91(4): 1608-1624, 2024 Apr.
Article en En | MEDLINE | ID: mdl-38102807
ABSTRACT

PURPOSE:

MP2RAGE parameter optimization is redefined to allow more time-efficient MR acquisitions, whereas the T1 -based synthetic imaging framework is used to obtain on-demand T1 -weighted contrasts. Our aim was to validate this concept on healthy volunteers and patients with multiple sclerosis, using plug-and-play parallel-transmission brain imaging at 7 T.

METHODS:

A "time-efficient" MP2RAGE sequence was designed with optimized parameters including TI and TR set as small as possible. Extended phase graph formalism was used to set flip-angle values to maximize the gray-to-white-matter contrast-to-noise ratio (CNR). Several synthetic contrasts (UNI, EDGE, FGATIR, FLAWSMIN , FLAWSHCO ) were generated online based on the acquired T1 maps. Experimental validation was performed on 4 healthy volunteers at various spatial resolutions. Clinical applicability was evaluated on 6 patients with multiple sclerosis, scanned with both time-efficient and conventional MP2RAGE parameterizations.

RESULTS:

The proposed time-efficient MP2RAGE protocols reduced acquisition time by 40%, 30%, and 19% for brain imaging at (1 mm)3 , (0.80 mm)3 and (0.65 mm)3 , respectively, when compared with conventional parameterizations. They also provided all synthetic contrasts and comparable contrast-to-noise ratio on UNI images. The flexibility in parameter selection allowed us to obtain a whole-brain (0.45 mm)3 acquisition in 19 min 56 s. On patients with multiple sclerosis, a (0.67 mm)3 time-efficient acquisition enhanced cortical lesion visualization compared with a conventional (0.80 mm)3 protocol, while decreasing the scan time by 15%.

CONCLUSION:

The proposed optimization, associated with T1 -based synthetic contrasts, enabled substantial decrease of the acquisition time or higher spatial resolution scans for a given time budget, while generating all typical brain contrasts derived from MP2RAGE.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Esclerosis Múltiple Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article País de afiliación: Francia Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Esclerosis Múltiple Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article País de afiliación: Francia Pais de publicación: Estados Unidos