Your browser doesn't support javascript.
loading
Preliminary study of cine-MRI compression in MR-guided radiotherapy.
Shang, Jiawen; Huang, Peng; Zhang, Ke; Hu, Zhihui; Yan, Hui.
Afiliación
  • Shang J; Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Huang P; Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Zhang K; Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Hu Z; Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Yan H; Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Quant Imaging Med Surg ; 13(12): 8009-8019, 2023 Dec 01.
Article en En | MEDLINE | ID: mdl-38106256
ABSTRACT

Background:

Cine-magnetic resonance imaging (MRI) is currently used in real-time tumor tracking during magnetic resonance (MR)-guided radiotherapy. As a type of MRI specified for motion tracking, a few minutes' acquisition results in thousands of 2-dimensional (2D) images. For MR-guided radiotherapy consisting of multiple treatment fractions, the large number of cine-MRI images would be disproportionate to the tight clinical data storage available. To alleviate this issue, the feasibility of compression of cine-MRI via video encoders was investigated in this study.

Methods:

The cine-MRI images were first sorted into 3 sequences according to their plane orientations. Then, each sequence was reordered according to their acquisition times [time-based (TB)] or content similarities [similarity-based (SB)]. As a result, 3 sequences were obtained for 3 plan orientations. Next, the obtained sequences were processed by a video encoder and the corresponding 3 video files were achieved. We employed 3 popular video encoders Motion JPEG (M-JPEG), Advanced Video Coding (AVC), and High Efficiency Video Coding (HEVC). The performances of the sequence reordering methods and video encoders were evaluated based on a total of 150 image sets.

Results:

The mean correlation quantities for SB sequences were higher than those for TB sequences by 3% (sagittal), 2% (coronal), and 1% (transverse), respectively. The average compression ratio (CR) yielded by the SB sequences was higher than that achieved by the TB sequences. Comparing with M-JPEG, the CRs obtained by AVC and HEVC were increased by 58% and 62% (sagittal), 16% and 23% (coronal), and 48% and 56% (transverse), respectively. Among the 3 video encoders, the highest CRs and restoration accuracy were achieved by HEVC.

Conclusions:

HEVC with inter-frame coding is more effective in reducing the redundant information in consecutive images. It is feasible to implement the video encoder for high-performance cine-MRI compression.