Trap-integrated fluorescence detection with silicon photomultipliers for sympathetic laser cooling in a cryogenic Penning trap.
Rev Sci Instrum
; 94(12)2023 Dec 01.
Article
en En
| MEDLINE
| ID: mdl-38109470
ABSTRACT
We present a fluorescence-detection system for laser-cooled 9Be+ ions based on silicon photomultipliers (SiPMs) operated at 4 K and integrated into our cryogenic 1.9 T multi-Penning-trap system. Our approach enables fluorescence detection in a hermetically sealed cryogenic Penning-trap chamber with limited optical access, where state-of-the-art detection using a telescope and photomultipliers at room temperature would be extremely difficult. We characterize the properties of the SiPM in a cryocooler at 4 K, where we measure a dark count rate below 1 s-1 and a detection efficiency of 2.5(3)%. We further discuss the design of our cryogenic fluorescence-detection trap and analyze the performance of our detection system by fluorescence spectroscopy of 9Be+ ion clouds during several runs of our sympathetic laser-cooling experiment.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
Rev Sci Instrum
Año:
2023
Tipo del documento:
Article
País de afiliación:
Alemania
Pais de publicación:
Estados Unidos