Clinical and structural outcomes of neocommissural alignment in transaxillary and transcarotid transcatheter aortic valve implantation with a self-expandable transcatheter heart valve.
Mosquera, Victor X; Muinelo-Paul, Adrian; Pato-López, Oscar; Bouzas-Mosquera, Alberto; Cuenca-Castillo, José J.
Afiliación
  • Mosquera VX; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Muinelo-Paul A; Instituto de Investigación Biomédica de A Coruña (INIBIC), University of A Coruña, A Coruña, Spain.
  • Pato-López O; University of Santiago de Compostela, Santiago de Compostela, Spain.
  • Bouzas-Mosquera A; Department of Cardiac Surgery, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
  • Cuenca-Castillo JJ; Department of Anesthesiology, Complejo Hospitalario Universitario de A Coruña, A Coruña, Spain.
JTCVS Tech ; 22: 150-158, 2023 Dec.
Article en En | MEDLINE | ID: mdl-38152173
ABSTRACT

Objectives:

This study analyzes neocommissural alignment and the clinical and hemodynamic outcomes after transaxillary and transcarotid implantation of the Acurate neo2 transcatheter heart valve.

Methods:

We performed a retrospective, single-center analysis of early outcomes after transaxillary and transcarotid implantation of the Acurate neo2 transcatheter heart valve. Primary outcomes were neocommisural alignment, in-hospital mortality, and valve hemodynamic performance. Commissural alignment between native and transcatheter heart valves was assessed by transesophageal echocardiogram before and after the procedure.

Results:

Between October 2021 and November 2022, 40 consecutive patients were treated with the Acurate neo2 through a transaxillary or transcarotid approach. Access was achieved via the left subclavian artery in 30 cases and the left common carotid artery in 10 cases, with a mean vessel diameter of 6.7 mm. Implants most commonly used were size M (37.5%), L (35%), and S (27.5%). On the basis of transesophageal echocardiogram analysis, there was no significant difference in mean commissural orientation between native (mean, 65.1°; SD, 41.3°) and neocommissures (mean, 64°; SD, 44.1°) (P = .661). Mean commissural orientation did not significantly differ between native and neocommissures (P = .661). Optimal alignment or mild commissural misalignment was achieved in 99.5% of cases. There were no cases of severe commissural misalignment. Postprocedural mean values for peak and mean gradients were 12.7 mm Hg and 5.2 mm Hg, respectively. There were 2 cases of moderate paravalvular leak and 4 cases of mild paravalvular leak.

Conclusions:

This patient-specific technique for transaxillary and transcarotid insertion of the Acurate neo2 delivery system prevents implantations with more than mild commissural misalignment and with a high device success rate.