Your browser doesn't support javascript.
loading
Effect of vitamin E-enhanced highly cross-linked polyethylene on wear rate and particle debris in anatomic total shoulder arthroplasty: a biomechanical comparison to ultrahigh-molecular-weight polyethylene.
Khan, Adam Z; Maxwell, Michael J; Parrott, Russell M; Bowler, Adam R; Glass, Evan A; Miller, Drew; Vasconcellos, Daniel; Brownhill, James R; Austin, Luke S; Cuff, Derek J; Murthi, Anand M; Smith, Matthew J; Wiater, J Michael; Jawa, Andrew.
Afiliación
  • Khan AZ; Northwest Permanente P.C., Portland, OR, USA.
  • Maxwell MJ; New England Baptist Hospital, Boston, MA, USA.
  • Parrott RM; Ignite Orthopedics, Winona Lake, IN, USA.
  • Bowler AR; New England Baptist Hospital, Boston, MA, USA.
  • Glass EA; New England Baptist Hospital, Boston, MA, USA.
  • Miller D; DePuy Synthes, Raynham, MA, USA.
  • Vasconcellos D; DePuy Synthes, Raynham, MA, USA.
  • Brownhill JR; DePuy Synthes, Raynham, MA, USA.
  • Austin LS; Department of Orthopaedic Surgery, The Rothman Institute, Philadelphia, PA, USA.
  • Cuff DJ; Suncoast Orthopaedic Surgery and Sports Medicine, Venice, FL, USA.
  • Murthi AM; MedStar Union Memorial Hospital, Baltimore, MD, USA.
  • Smith MJ; Department of Orthopedic Surgery, University of Missouri, Columbia, MO, USA.
  • Wiater JM; Department of Orthopaedic Surgery, Beaumont Hospital, Royal Oak, MI, USA.
  • Jawa A; New England Baptist Hospital, Boston, MA, USA. Electronic address: andrewjawa@gmail.com.
J Shoulder Elbow Surg ; 33(7): 1465-1472, 2024 Jul.
Article en En | MEDLINE | ID: mdl-38182025
ABSTRACT

BACKGROUND:

Particle-induced osteolysis resulting from polyethylene wear remains a source of implant failure in anatomic total shoulder designs. Modern polyethylene components are irradiated in an oxygen-free environment to induce cross-linking, but reducing the resulting free radicals with melting or heat annealing can compromise the component's mechanical properties. Vitamin E has been introduced as an adjuvant to thermal treatments. Anatomic shoulder arthroplasty models with a ceramic head component have demonstrated that vitamin E-enhanced polyethylene show improved wear compared with highly cross-linked polyethylene (HXLPE). This study aimed to assess the biomechanical wear properties and particle size characteristics of a novel vitamin E-enhanced highly cross-linked polyethylene (VEXPE) glenoid compared to a conventional ultrahigh-molecular-weight polyethylene (UHMWPE) glenoid against a cobalt chromium molybdenum (CoCrMo) head component.

METHODS:

Biomechanical wear testing was performed to compare the VEXPE glenoid to UHMWPE glenoid with regard to pristine polyethylene wear and abrasive endurance against a polished CoCrMo alloy humeral head in an anatomic shoulder wear-simulation model. Cumulative mass loss (milligrams) was recorded, and wear rate calculated (milligrams per megacycle [Mc]). Under pristine wear conditions, particle analysis was performed, and functional biologic activity (FBA) was calculated to estimate particle debris osteolytic potential. In addition, 95% confidence intervals for all testing conditions were calculated.

RESULTS:

The average pristine wear rate was statistically significantly lower for the VEXPE glenoid compared with the HXLPE glenoid (0.81 ± 0.64 mg/Mc vs. 7.00 ± 0.45 mg/Mc) (P < .05). Under abrasive wear conditions, the VEXPE glenoid had a statistically significant lower average wear rate compared with the UHMWPE glenoid comparator device (18.93 ± 5.80 mg/Mc vs. 40.47 ± 2.63 mg/Mc) (P < .05). The VEXPE glenoid demonstrated a statistically significant improvement in FBA compared with the HXLPE glenoid (0.21 ± 0.21 vs. 1.54 ± 0.49 (P < .05).

CONCLUSIONS:

A new anatomic glenoid component with VEXPE demonstrated significantly improved pristine and abrasive wear properties with lower osteolytic particle debris potential compared with a conventional UHMWPE glenoid component. Vitamin E-enhanced polyethylene shows early promise in shoulder arthroplasty components. Long-term clinical and radiographic investigation needs to be performed to verify if these biomechanical wear properties translate to diminished long-term wear, osteolysis, and loosening.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polietilenos / Diseño de Prótesis / Vitamina E / Ensayo de Materiales / Falla de Prótesis / Artroplastía de Reemplazo de Hombro / Prótesis de Hombro Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Shoulder Elbow Surg Asunto de la revista: ORTOPEDIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polietilenos / Diseño de Prótesis / Vitamina E / Ensayo de Materiales / Falla de Prótesis / Artroplastía de Reemplazo de Hombro / Prótesis de Hombro Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Shoulder Elbow Surg Asunto de la revista: ORTOPEDIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos