Comparison of Evolut-R 34 mm Valve and Smaller Evolut-R Valves in Patients Undergoing Transcatheter Aortic Valve Implantation and Determination of Mild Paravalvular Leak Predictors.
Kiliç, Raif; Güzel, Tuncay; Aktan, Adem; Demir, Muhammed; Günlü, Serhat; Arslan, Bayram; Ertas, Faruk.
Afiliación
  • Kiliç R; Department of Cardiology, Çermik State Hospital, Diyarbakir, Türkiye.
  • Güzel T; Department of Cardiology, Health Science University, Gazi Yasargil Training and Research Hospital, Diyarbakir, Türkiye.
  • Aktan A; Department of Cardiology, Mardin Training and Research Hospital, Mardin, Türkiye.
  • Demir M; Department of Cardiology, Dicle University Faculty of Medicine, Diyarbakir, Türkiye.
  • Günlü S; Department of Cardiology, Mardin Artuklu University Faculty of Medicine, Mardin, Türkiye.
  • Arslan B; Department of Cardiology, Ergani State Hospital, Diyarbakir, Türkiye.
  • Ertas F; Department of Cardiology, Dicle University Faculty of Medicine, Diyarbakir, Türkiye.
Anatol J Cardiol ; 2024 Jan 15.
Article en En | MEDLINE | ID: mdl-38221789
ABSTRACT

OBJECTIVE:

The main purpose of this study was to evaluate and compare the in-hospital, 1-month and 1-year post-procedure outcomes of patients treated with Evolut-R 34 mm and Evolut-R 23/26/29 mm devices. Additionally, the study aimed to identify factors that could predict the occurrence of ≥ mild paravalvular leaks (PVL).

METHODS:

Between April 2015 and May 2022, 269 consecutive patients who underwent transcatheter aortic valve implantation (TAVI) with Evolut-R 34 mm (n = 66, 24.5%) and Evolut-R 23/26/29 mm (n = 203, 75.5%) devices in a single center were retrospectively analyzed.

RESULTS:

Patients in the Evolut-R 34 mm group had a lower female sex ratio (16.7% vs. 66.5%, P <.001, respectively), ejection fraction (50.7 ± 10.1% vs. 54.5 ± 9.3%, P =.016, respectively), and mean aortic gradient (7.4 ± 3.3 vs. 9.2 ± 5.0, P =.026, respectively) compared to the Evolut-R 23/26/29 mm group. The groups did not exhibit any statistically significant distinctions with regard to technical success, the need for a permanent pacemaker, occurrences of stroke, major vascular complications, PVL, major adverse cardiovascular and cerebrovascular events, or mortality. Peak velocity was confirmed as a significant predictor of ≥mild PVL in both patient groups in the receiver operating characteristic curve analysis. In logistic regression analysis; In patients with Evolut-R 34 mm valve, pre-TAVI aortic valve peak velocity (odds ratio (OR) = 23.202; P =.019) and calcium volume 800 Hounsfield Units (mm3) (OR = 1.017; P <.001) were independent predictors of ≥mild PVL.

CONCLUSION:

The Evolut-R 34 mm valve has shown comparable in-hospital results with smaller valve sizes. Pre-TAVI aortic valve peak velocity and calcium volume predicted ≥ mild PVL in Evolut-R 34 mm patients.