Your browser doesn't support javascript.
loading
Mind the Gap in TAVR: Recognizing and Managing Misloaded Self-Expanding TAVR Devices.
Omar, Aninda; Damlin, Anna; Rück, Andreas; Settergren, Magnus; Verouhis, Dinos; Linder, Rickard; Meduri, Christopher U; Saleh, Nawzad.
Afiliación
  • Omar A; Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden.
  • Damlin A; Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden.
  • Rück A; Unit of Cardiology, Department of Medicine, Karolinska Institutet, Stockholm, Sweden.
  • Settergren M; Department of Clinical Physiology, Karolinska University Hospital, Stockholm, Sweden.
  • Verouhis D; Department of Molecular Medicine and Surgery, Karolinska Institutet, Stockholm, Sweden.
  • Linder R; Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden.
  • Meduri CU; Unit of Cardiology, Department of Medicine, Karolinska Institutet, Stockholm, Sweden.
  • Saleh N; Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden.
JACC Case Rep ; 29(3): 102192, 2024 Feb 07.
Article en En | MEDLINE | ID: mdl-38361571
ABSTRACT
Misloading during transcatheter aortic valve replacement (TAVR) is rare but can cause unpredictable valve release if unrecognized. We describe how to identify a misloaded ACURATE neo2 device, and 3 methods to solve this by using a modified technique of valve deployment, ipsilateral extraction, and contralateral valve externalization with extracorporeal valve release.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: JACC Case Rep Año: 2024 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Países Bajos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: JACC Case Rep Año: 2024 Tipo del documento: Article País de afiliación: Suecia Pais de publicación: Países Bajos