Mind the Gap in TAVR: Recognizing and Managing Misloaded Self-Expanding TAVR Devices.
JACC Case Rep
; 29(3): 102192, 2024 Feb 07.
Article
en En
| MEDLINE
| ID: mdl-38361571
ABSTRACT
Misloading during transcatheter aortic valve replacement (TAVR) is rare but can cause unpredictable valve release if unrecognized. We describe how to identify a misloaded ACURATE neo2 device, and 3 methods to solve this by using a modified technique of valve deployment, ipsilateral extraction, and contralateral valve externalization with extracorporeal valve release.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Tipo de estudio:
Prognostic_studies
Idioma:
En
Revista:
JACC Case Rep
Año:
2024
Tipo del documento:
Article
País de afiliación:
Suecia
Pais de publicación:
Países Bajos