In Vitro Head-to-Head Comparison of Flow Reduction between Fibered and Non-Fibered Pushable Coils.
Yoon, Jong-Tae; Kwon, Boseong; Choi, Joon Ho; Hwang, Sun Moon; Kim, Mihyeon; Hwang, Sungbin; Song, Yunsun; Lee, Deok Hee.
Afiliación
  • Yoon JT; Department of Radiology, AMIST, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Kwon B; Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Choi JH; Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Hwang SM; Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Kim M; Department of Radiology, AMIST, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Hwang S; Department of Radiology, AMIST, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Song Y; Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Lee DH; Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Neurointervention ; 19(1): 31-38, 2024 Mar.
Article en En | MEDLINE | ID: mdl-38373422
ABSTRACT

PURPOSE:

To compare the embolization effects of a non-fibered pushable coil with a conventional fibered pushable coil in an in vitro bench-top experiment. MATERIALS AND

METHODS:

A simplified vascular phantom with 4 channels (1 for the non-fibered coil, 1 for the fibered coil, and 2 for continuous circuit flow) was used. A single coil of the longest length was inserted to evaluate the effect of single-coil embolization, and 3 consecutive coils were inserted to assess the effect of multiple-coil embolization. Post-embolization angiography was performed to obtain flow variables (time to peak [TTP], relative peak intensity [rPI], and angiographic flow reduction score [AFRS]) from time density curves. The packing densities of the two coil types were calculated, and the AFRS of each channel was determined by dividing the TTP by the rPI.

RESULTS:

When inserting a single coil, the conventional fibered coil demonstrated better flow reduction, as indicated by a higher AFRS (25.6 vs. 17.4, P=0.034). However, the non-fibered coil exhibited a significantly higher packing density (12.9 vs. 2.4, P=0.001). Similar trends were observed with multiple coils.

CONCLUSION:

The conventional fibered pushable coil showed better flow reduction efficiency, while the non-fibered pushable coil had a higher packing density, likely due to the flexibility of the coil loops. A better understanding of the distinct characteristics of different pushable coils can enhance the outcomes of various vascular embolization.