Your browser doesn't support javascript.
loading
Qualitative and quantitative image quality of coronary CT angiography using photon-counting computed tomography: Standard and Ultra-high resolution protocols.
Vattay, Borbála; Boussoussou, Melinda; Vecsey-Nagy, Milán; Kolossváry, Márton; Juhász, Dénes; Kerkovits, Nóra; Balogh, Hanna; Nagy, Norbert; Vértes, Miklós; Kiss, Máté; Kubovje, Anikó; Merkely, Béla; Maurovich Horvat, Pál; Szilveszter, Bálint.
Afiliación
  • Vattay B; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
  • Boussoussou M; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
  • Vecsey-Nagy M; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
  • Kolossváry M; Gottsegen National Cardiovascular Center, 29 Haller Utca, 1096, Budapest, Hungary; Physiological Controls Research Center, University Research and Innovation Center, Óbuda University, Bécsi Ut 96/B, 1034, Budapest, Hungary.
  • Juhász D; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
  • Kerkovits N; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Balogh H; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Nagy N; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Vértes M; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Kiss M; Siemens Healthcare GmbH, Forchheim, Germany.
  • Kubovje A; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Merkely B; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
  • Maurovich Horvat P; Medical Imaging Center, Semmelweis University, 1082 Budapest, Korányi Sándor Street 2., Hungary.
  • Szilveszter B; Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary. Electronic address: szilveszter.balint@med.semmelweis-univ.hu.
Eur J Radiol ; 175: 111426, 2024 Jun.
Article en En | MEDLINE | ID: mdl-38493558
ABSTRACT

PURPOSE:

We aimed to identify the optimal reconstruction settings based on qualitative and quantitative image quality parameters on standard and ultra-high resolution (UHR) images using photon-counting CT (PCCT).

METHOD:

We analysed 45 patients, 29 with standard and 16 with UHR acquisition, applying both smoother and sharper kernel settings. Coronary CT angiography images were performed on a dual-source PCCT system using standard (0.4/0.6 mm slice thickness, Bv40/Bv44 kernels, QIR levels 0-4) or UHR acquisition (0.2/0.4 mm slice thickness, Bv44/Bv56 kernels, QIR levels 0-4). Qualitative image quality was assessed using a 4-point Likert scale. Image noise (SD), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated in both the proximal and distal segments.

RESULTS:

On standard resolution, larger slice thickness resulted in an average increase of 12.5 % in CNR, whereas sharper kernel led to an average 8.7 % decrease in CNR. Highest CNR was measured on 0.6 mm, Bv40, QIR4 images and lowest on 0.4 mm, Bv44, QIR0 images 25.8 ± 4.1vs.8.3 ± 1.6 (p < 0.001). On UHR images, highest CNR was observed on 0.4 mm, Bv40, QIR4 and lowest on 0.2 mm, Bv56 and QIR0 images 21.5 ± 3.9vs.3.6 ± 0.8 (p < 0.001). Highest qualitative image quality was found on images with Bv44 kernel and QIR level 3/4 with both slice thicknesses on standard reconstruction. Additionally, Bv56 with QIR4 on 0.2 mm slice thickness images showed highest subjective image quality. Preserved distal vessel visualization was detected using QIR 2-4, Bv56 and 0.2 mm slice thickness.

CONCLUSIONS:

Photon-counting CT demonstrated high qualitative and quantitative image quality for the assessment of coronaries and stents.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Angiografía Coronaria / Fotones / Angiografía por Tomografía Computarizada Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Radiol Año: 2024 Tipo del documento: Article País de afiliación: Hungria

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Angiografía Coronaria / Fotones / Angiografía por Tomografía Computarizada Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Radiol Año: 2024 Tipo del documento: Article País de afiliación: Hungria