Your browser doesn't support javascript.
loading
New robotic platform for transoral robotic surgery: an IDEAL stage 0 study.
Arora, Asit; Faulkner, Jack; Paleri, Vinidh; Kapoor, Karan; Al-Lami, Ali; Olaleye, Oladejo; Winter, Stuart; Oikonomou, Geogios; Ofo, Enyi; Ourselin, Sebastien; Dasgupta, Prokar; Slack, Mark; Jeannon, Jean-Pierre.
Afiliación
  • Arora A; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Faulkner J; Head and Neck Surgery, Guy's Hospital, London, UK.
  • Paleri V; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Kapoor K; Head and Neck Surgery, Guy's Hospital, London, UK.
  • Al-Lami A; Royal Marsden Hospital NHS Trust, London, UK.
  • Olaleye O; Surrey and Sussex Healthcare NHS Trust, Redhill, Surrey, UK.
  • Winter S; East Kent Hospitals University NHS Foundation Trust, Canterbury, Kent, UK.
  • Oikonomou G; Otolaryngology, Head and Neck Surgery, University Hospitals of Leicester NHS Trust, Leicester, UK.
  • Ofo E; Department of Surgical Sciences, University of Oxford Nuffield, Oxford, Oxfordshire, UK.
  • Ourselin S; St George's University Hospitals NHS Foundation Trust, London, UK.
  • Dasgupta P; St George's University Hospitals NHS Foundation Trust, London, UK.
  • Slack M; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Jeannon JP; MRC Centre for Transplantation, NIHR Biomedical Research Centre, King's College, London, UK.
BMJ Surg Interv Health Technol ; 6(1): e000181, 2024.
Article en En | MEDLINE | ID: mdl-38500710
ABSTRACT

Objectives:

This study aims to assess the feasibility to perform transoral robotic surgery (TORS) with a new robotic platform, the Versius Surgical System (CMR Surgical, UK) in a preclinical cadaveric setting in accordance to stage 0 of the IDEAL-D framework.

Design:

IDEAL stage 0 preclinical assessment of the Versius Robotic System in TORS in human cadavers.

Setting:

All procedures were performed in a simulated operating theatre environment at a UK surgical training centre.

Participants:

11 consultant head and neck surgeons from the UK, mainland Europe and the USA took part in TORS procedures on six human cadavers.

Interventions:

3 key index procedures were assessed that represent the core surgical workload of TORS lateral oropharyngectomy, tongue base resection and partial supraglottic laryngectomy. Main outcome

measures:

The primary outcome was the successful completion of each surgical procedure. Secondary outcomes included the optimisation of system setup, instrumentation and surgeon-reported outcomes for feasibility of each component procedural step.

Results:

33 cadaveric procedures were performed and 32 were successfully completed. One supraglottic laryngectomy was not fully completed due to issues dividing the epiglottic cartilage with available instrumentation. Surgeon-reported outcomes met the minimal level of feasibility in all procedures and a consensus that it is feasible to perform TORS with Versius was reached. Available instrumentation was not representative of other robotic platforms used in TORS and further instrument optimisation is recommended before wider dissemination.

Conclusions:

It is feasible to perform TORS with the Versius Surgical System (CMR Surgical) within a pre-clinical cadaveric setting. Clinical evaluation is needed and appropriate with the system. Further instrument development and optimisation is desirable.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: BMJ Surg Interv Health Technol Año: 2024 Tipo del documento: Article Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: BMJ Surg Interv Health Technol Año: 2024 Tipo del documento: Article Pais de publicación: Reino Unido