Skin marker combined with surface-guided auto-positioning for breast DIBH radiotherapy daily initial patient setup: An optimal schedule for both accuracy and efficiency.
Lai, Jianjun; Luo, Zhizeng; Jiang, Lu; Hu, Haili; Gao, Chang; Zhang, Chuanfeng; Chen, Liting; Wu, Jing; Wu, Zhibing.
Afiliación
  • Lai J; Instiute of Intelligent Control and Robotics, Hangzhou Dianzi University, Hangzhou, China.
  • Luo Z; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Jiang L; Instiute of Intelligent Control and Robotics, Hangzhou Dianzi University, Hangzhou, China.
  • Hu H; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Gao C; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Zhang C; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Chen L; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Wu J; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
  • Wu Z; Department of Radiation Oncology, Zhejiang Hospital, Hangzhou, China.
J Appl Clin Med Phys ; 25(7): e14319, 2024 Jul.
Article en En | MEDLINE | ID: mdl-38522035
ABSTRACT
BACKGROUND AND

PURPOSE:

By employing three surface-guided radiotherapy (SGRT)-assisted positioning methods, we conducted a prospective study of patients undergoing SGRT-based deep inspiration breath-hold (DIBH) radiotherapy using a Sentine/Catalys system. The aim of this study was to optimize the initial positioning workflow of SGRT-DIBH radiotherapy for breast cancer. MATERIALS AND

METHODS:

A total of 124 patients were divided into three groups to conduct a prospective comparative study of the setup accuracy and efficiency for the daily initial setup of SGRT-DIBH breast radiotherapy. Group A was subjected to skin marker plus SGRT verification, Group B underwent SGRT optical feedback plus auto-positioning, and Group C was subjected to skin marker plus SGRT auto-positioning. We evaluated setup accuracy and efficiency using cone-beam computed tomography (CBCT) verification data and the total setup time.

RESULTS:

In groups A, B, and C, the mean and standard deviation of the translational setup-error vectors were small, with the highest values of the three directions observed in group A (2.4 ± 1.6, 2.9 ± 1.8, and 2.8 ± 2.1 mm). The rotational vectors in group B (1.8 ± 0.7°, 2.1 ± 0.8°, and 1.8 ± 0.7°) were significantly larger than those in groups A and C, and the Group C setup required the shortest amount of time, at 1.5 ± 0.3 min, while that of Group B took the longest time, at 2.6 ± 0.9 min.

CONCLUSION:

SGRT one-key calibration was found to be more suitable when followed by skin marker/tattoo and in-room laser positioning, establishing it as an optimal daily initial set-up protocol for breast DIBH radiotherapy. This modality also proved to be suitable for free-breathing breast cancer radiotherapy, and its widespread clinical use is recommended.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosificación Radioterapéutica / Planificación de la Radioterapia Asistida por Computador / Neoplasias de la Mama / Radioterapia de Intensidad Modulada / Tomografía Computarizada de Haz Cónico / Posicionamiento del Paciente / Contencion de la Respiración Límite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosificación Radioterapéutica / Planificación de la Radioterapia Asistida por Computador / Neoplasias de la Mama / Radioterapia de Intensidad Modulada / Tomografía Computarizada de Haz Cónico / Posicionamiento del Paciente / Contencion de la Respiración Límite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2024 Tipo del documento: Article País de afiliación: China