Enhanced Myocardial Tissue Visualization: A Comparative Cardiovascular Magnetic Resonance Study of Gradient-Spin Echo-STIR and Conventional STIR Imaging.
Dehghani, Sadegh; Shirani, Shapoor; Jazayeri Gharebagh, Elahe.
Afiliación
  • Dehghani S; Radiation Sciences Department, School of Allied Medical Sciences, Tehran University of Medical Sciences, Tehran, Iran.
  • Shirani S; Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran.
  • Jazayeri Gharebagh E; Radiation Sciences Department, School of Allied Medical Sciences, Tehran University of Medical Sciences, Tehran, Iran.
Int J Biomed Imaging ; 2024: 8456669, 2024.
Article en En | MEDLINE | ID: mdl-38590625
ABSTRACT

Purpose:

This study is aimed at evaluating the efficacy of the gradient-spin echo- (GraSE-) based short tau inversion recovery (STIR) sequence (GraSE-STIR) in cardiovascular magnetic resonance (CMR) imaging compared to the conventional turbo spin echo- (TSE-) based STIR sequence, specifically focusing on image quality, specific absorption rate (SAR), and image acquisition time.

Methods:

In a prospective study, we examined forty-four normal volunteers and seventeen patients referred for CMR imaging using conventional STIR and GraSE-STIR techniques. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), image quality, T2 signal intensity (SI) ratio, SAR, and image acquisition time were compared between both sequences.

Results:

GraSE-STIR showed significant improvements in image quality (4.15 ± 0.8 vs. 3.34 ± 0.9, p = 0.024) and cardiac motion artifact reduction (7 vs. 18 out of 53, p = 0.038) compared to conventional STIR. Furthermore, the acquisition time (27.17 ± 3.53 vs. 36.9 ± 4.08 seconds, p = 0.041) and the local torso SAR (<13% vs. <17%, p = 0.047) were significantly lower for GraSE-STIR compared to conventional STIR in short-axis plan. However, no significant differences were shown in T2 SI ratio (p = 0.141), SNR (p = 0.093), CNR (p = 0.068), and SAR (p = 0.071) between these two sequences.

Conclusions:

GraSE-STIR offers notable advantages over conventional STIR sequence, with improved image quality, reduced motion artifacts, and shorter acquisition times. These findings highlight the potential of GraSE-STIR as a valuable technique for routine clinical CMR imaging.