Performance evaluation of a newly developed 2019-nCoV nucleic acid detection kit based on Ion Proton sequencing platform and its comparison with the MGI Tech (DNBSEQ-G99) platform.
He, Yuting; Yang, Tingting; Zhong, Guosheng; Yu, Xuegao; Zhao, Zhiwei; Shi, Yaling; Huang, Bin.
Afiliación
  • He Y; Department of Clinical Laboratory, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China.
  • Yang T; Department of Clinical Laboratory, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China.
  • Zhong G; Department of Clinical Laboratory, The Sixth Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China.
  • Yu X; Department of Clinical Laboratory, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China.
  • Zhao Z; Department of Clinical Laboratory, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China.
  • Shi Y; Department of Clinical Laboratory, Guangzhou Eighth People's Hospital, Guangzhou Medical University.
  • Huang B; Department of Clinical Laboratory, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, PR China. Electronic address: huangb3@mail.sysu.edu.cn.
Diagn Microbiol Infect Dis ; 109(3): 116323, 2024 Jul.
Article en En | MEDLINE | ID: mdl-38703530
ABSTRACT

PURPOSE:

To evaluate the performance of a newly developed 2019-nCoV nucleic acid detection kit based on Ion Proton sequencing platform and make comparation with MGI Tech (DNBSEQ-G99) platform.

METHODS:

References and clinical samples were used to evaluate the precision, agreement rate, limit of detection (LOD), anti-interference ability and analytical specificity. Twenty-seven clinical specimens were used to make comparison between two platforms.

RESULTS:

The kit showed good intra-assay, inter-assay, inter-day precision between different operators and laboratories, fine agreement rate with references, a relatively low LOD of 1 × 103 copies/ml, anti-interference capability of 5 % whole blood and 1mg/ml mucin and no cross reaction with twenty-nine common clinical pathogens. Consistency of variant classification was observed between two platforms. The WGS from Ion Proton tended to have higher coverage and less missing data.

CONCLUSIONS:

The newly developed kit has shown satisfactory performances and excellent consistency with DNBSEQ-G99, making it a good alternative choice clinically.