Impact of beam complexity on plan delivery accuracy verification of a transmission detector in volumetric modulated arc therapy.
Radici, Lorenzo; Petrucci, Edoardo; Casanova Borca, Valeria; Cante, Domenico; Piva, Cristina; Pasquino, Massimo.
Afiliación
  • Radici L; Medical Physics Department, ASL TO4, Ivrea, Italy.
  • Petrucci E; Medical Physics Department, ASL TO4, Ivrea, Italy.
  • Casanova Borca V; Medical Physics Department, ASL TO4, Ivrea, Italy.
  • Cante D; Radiotherapy Department, ASL TO4, Ivrea, Italy.
  • Piva C; Radiotherapy Department, ASL TO4, Ivrea, Italy.
  • Pasquino M; Medical Physics Department, ASL TO4, Ivrea, Italy. Electronic address: mpasquino@aslto4.piemonte.it.
Phys Med ; 122: 103387, 2024 Jun.
Article en En | MEDLINE | ID: mdl-38797025
ABSTRACT

OBJECTIVE:

To study the effect of beam complexity on VMAT delivery accuracy evaluated by means of a transmission detector, together with the possibility of scoring plan complexity.

METHODS:

43 clinical VMAT plans delivered by a TrueBeam linear accelerator to both Delta4 Discover and Delta4 Phantom+ for patient-specific quality assurance were evaluated. Global Dose-γ analysis, MLC-γ analysis, percentage of leaves with a deviation between planned and measured leaf tip position lower than 1 mm (LD) were computed. Modulation complexity score (MCSv), average leaf travel (LT), a multiplicative combination of LT and MCSv (LTMCS), percentage of leaves with speed lower than 5 mm/s (LS), from 5 to 20 mm/s (MS), higher than 20 mm/s (HS) and the average value of leaf speed (MLCSav) were evaluated by means of an home-made Matlab script.

RESULTS:

Dose-γ passing rate showed a moderate correlation with MCSv, LT, MLCSav, LS and HS, while a stronger positive correlation was found with LTMCS. A strong correlation was observed between LD and both LT and leaves speed, while a weak correlation was observed with MCSv. A correlation between MLC-γ pass rate and plan complexity parameters was found except for MCSv; a moderate correlation with LS was observed, while all other parameters showed weak correlations.

CONCLUSIONS:

The study confirmed the possibility to establish correlations between plan complexity indices versus dose distribution and MLC parameters measured by a transmissive detector. Further investigation is necessary to define specific values of the complexity indices to evaluate whether a VMAT plan is deliverable as intended.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2024 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Italia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2024 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Italia