Your browser doesn't support javascript.
loading
Multiangle Light Scattering as a Lentivirus Purification Process Analytical Technology.
Sripada, Sobhana A; Barbieri, Eduardo; Shastry, Shriarjun; Wuestenhagen, Elena; Aldinger, Annika; Rammo, Oliver; Schulte, Michael M; Daniele, Michael; Menegatti, Stefano.
Afiliación
  • Sripada SA; Department of Chemical and Biomolecular Engineering, NC State University, 911 Partners Way, Raleigh, North Carolina 27606, United States.
  • Barbieri E; NC-VVIRAL, NC State University, 1840 Entrepreneur Dr, Raleigh, North Carolina 27606, United States.
  • Shastry S; Department of Chemical and Biomolecular Engineering, NC State University, 911 Partners Way, Raleigh, North Carolina 27606, United States.
  • Wuestenhagen E; LigaTrap Technologies LLC, 1791 Varsity Drive, Suite #150, Raleigh, North Carolina 27606, United States.
  • Aldinger A; Department of Chemical and Biomolecular Engineering, NC State University, 911 Partners Way, Raleigh, North Carolina 27606, United States.
  • Rammo O; Biomanufacturing Training and Education Center, NC State University, 850 Oval Dr, Raleigh, North Carolina 27606, United States.
  • Schulte MM; NC-VVIRAL, NC State University, 1840 Entrepreneur Dr, Raleigh, North Carolina 27606, United States.
  • Daniele M; Merck KGaA, Darmstadt 64293, Germany.
  • Menegatti S; Merck KGaA, Darmstadt 64293, Germany.
Anal Chem ; 96(23): 9593-9600, 2024 06 11.
Article en En | MEDLINE | ID: mdl-38804040
ABSTRACT
The limited biomolecular and functional stability of lentiviral vectors (LVVs) for cell therapy poses the need for analytical tools that can monitor their titers and activity throughout the various steps of expression and purification. In this study, we describe a rapid (25 min) and reproducible (coefficient of variance ∼0.5-2%) method that leverages size exclusion chromatography coupled with multiangle light scattering detection (SEC-MALS) to determine size, purity, and particle count of LVVs purified from bioreactor harvests. The SEC-MALS data were corroborated by orthogonal methods, namely, dynamic light scattering (DLS) and transmission electron microscopy. The method was also evaluated for robustness in the range of 2.78 × 105-2.67 × 107 particles per sample. Notably, MALS-based particle counts correlated with the titer of infectious LVVs measured via transduction assays (R2 = 0.77). Using a combination of SEC-MALS and DLS, we discerned the effects of purification parameters on LVV quality, such as the separation between heterogeneous LV, which can facilitate critical decision-making in the biomanufacturing of gene and cell therapies.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Lentivirus / Dispersión Dinámica de Luz Límite: Humans Idioma: En Revista: Anal Chem Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Lentivirus / Dispersión Dinámica de Luz Límite: Humans Idioma: En Revista: Anal Chem Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos