Comparison of the HugeMed video laryngoscope with the Macintosh direct laryngoscope for nasotracheal tracheal intubation in children undergoing dental treatment: a randomized controlled clinical study.
Expert Rev Med Devices
; : 1-8, 2024 Jun 03.
Article
en En
| MEDLINE
| ID: mdl-38829609
ABSTRACT
OBJECTIVES:
The aim of this study was to compare the performance of the HugeMed® videolaryngoscope with a direct Macintosh laryngoscope for nasotracheal intubation.METHODS:
Eighty-eight children aged 4-10 years were randomly assigned to either the HugeMed® videolaryngoscope (HVL) or the Macintosh direct laryngoscope (MDL) group. Intubation difficulty, glottic view grade, time-to-intubation, number of tracheal intubation attempts, use of external laryngeal manipulation and Magill forceps, recovery time, pediatric-anesthesia-delirium-scale (PAEDS) scores, pain due to tracheal intubation, and laryngeal bleeding were evaluated.RESULTS:
Easy tracheal intubation incidence was higher in the HVL group than that in the MDL group (p = 0.001). Glottic view was better in the HVL group as compared to the MDL group (p = 0.027). There was no difference between the groups in terms of time-to-tracheal intubation, number of tracheal intubation attempts, Magill forceps usage, pain, and bleeding due to tracheal intubation. The need for external laryngeal manipulation (p = 0.004) and PAEDS scores (p = 0.006) were higher in the MDL group than those in the HVL group.CONCLUSION:
HugeMed® videolaryngoscope may provide easier tracheal intubation, create a better glottic view, and significantly reduce the need for additional manipulation compared to the Macintosh direct laryngoscope, for nasotracheal intubation. CLINICAL TRIAL REGISTRATION www.clinicaltrial.gov identifier is NCT05121597.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
Expert Rev Med Devices
Asunto de la revista:
DIAGNOSTICO POR IMAGEM
/
TERAPEUTICA
Año:
2024
Tipo del documento:
Article
País de afiliación:
Turquía