Correlation of plasmatic sodium determined by the laboratory and that determined by the dialysis machine.
Casals, Joaquim; Broseta, José Jesús; Fernández, Rosa María; Rodriguez-Espinosa, Diana; Del Risco, Jimena; Gómez, Miquel; Rodas, Lida M; Arias-Guillén, Marta; Vera, Manel; Fontseré, Néstor; Rico, Naira; Maduell, Francisco.
Afiliación
  • Casals J; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Broseta JJ; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain. Electronic address: jjbroseta@clinic.cat.
  • Fernández RM; Centro de Diagnóstico Biomédico, Hospital Clínic, Barcelona, Spain.
  • Rodriguez-Espinosa D; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Del Risco J; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Gómez M; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Rodas LM; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Arias-Guillén M; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Vera M; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Fontseré N; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
  • Rico N; Centro de Diagnóstico Biomédico, Hospital Clínic, Barcelona, Spain.
  • Maduell F; Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
Nefrologia (Engl Ed) ; 44(3): 417-422, 2024.
Article en En | MEDLINE | ID: mdl-38890062
ABSTRACT

INTRODUCTION:

Changes in plasma sodium concentration (pNa, expressed in mEq/L) are common in hemodialysis (HD) patients. Hemodialysis monitors can estimate pNa by using an internal algorithm based on ion dialysance measurements. The present study studies the accuracy of the correlation between the pNa estimated by the dialysis monitor and that measured by the biochemistry laboratory at our center. MATERIAL AND

METHODS:

A single-centre prospective observational study in patients on a chronic HD program with the 6008 CAREsystem monitor and standard sodium (138mmol/L) and bicarbonate (32mmol/L) prescriptions. Venous blood samples were drawn from each patient before and after each HD session to ensure inter- and intra-individual validity. The pNa was measured in the biochemistry laboratory using indirect potentiometry and simultaneously the estimated pNa by the HD monitor was recorded at the beginning and at the end of the HD session. For statistical analysis, a scatterplot was made, and Spearman's correlation quotient was calculated. In addition, the differences between both methods were represented as Bland-Altman diagrams.

RESULTS:

The pre-dialysis pNa measured in the laboratory was 137.49±3.3, and that of the monitor, 137.96±2.91, with a correlation with R2 value of 0.683 (p<0.001). The post-dialysis pNa measured in the laboratory was 137.08±2.23, and that of the monitor was 138.87±1.88, with an R2 of 0.442 (p<0.001). On the Bland-Altman plots, the pre-dialysis pNa has a systematic error of 0.49, in favor of the monitor-estimated pNa, with a 95% confidence interval (CI) of (-3.24 to a 4.22). In the post-dialysis pNa, a systematic error of 1.79 with a 95% CI of (-1.64 to 5.22) was obtained.

CONCLUSION:

The correlation between the pNa estimated by Fresnius 6008 CAREsystem HD monitor and that measured by the laboratory is good, especially pre-dialysis measurements. Further studies should verify the external validity of these results.