Development and validation of a non-invasive method for quantifying amino acids in human saliva.
Hasan, Md Mehedi; Razu, Mamudul Hasan; Akter, Sonia; Mou, Salma Akter; Islam, Minhazul; Khan, Mala.
Afiliación
  • Hasan MM; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
  • Razu MH; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
  • Akter S; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
  • Mou SA; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
  • Islam M; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
  • Khan M; Bangladesh Reference Institute for Chemical Measurements Dhaka Bangladesh dg.mic@bricm.gov.bd mhrazu@bricm.gov.bd.
RSC Adv ; 14(31): 22292-22303, 2024 Jul 12.
Article en En | MEDLINE | ID: mdl-39010921
ABSTRACT
As an analytical matrix, saliva has superior characteristics than blood and urine. Saliva collection is, first and foremost, non-invasive, making it convenient, painless, and secure for more susceptible people. Second, it does not need professional training for medical personnel, resulting in cost-effectiveness and suitability for extensive collection in support of research. In this study, we developed a method and used it to quantify 13 salivary-free amino acid (SFAA) profiles to support the early clinical diagnosis of diseases using LC-MS/MS. Using an Intrada Amino Acid column (100 × 3 mm, 3 µm), chromatographic separation was accomplished with a binary gradient elution, and an electrospray ionisation source running in the positive ionisation mode was chosen for data collection using the multiple reaction monitoring (MRM) modes. Amino acids were extracted from saliva using acetonitrile. In the MRM mode, LODs and LOQs for ten amino acids were in the range of 0.06-2.50 µM and 0.19-7.58 µM, respectively, and those values were in the range of 1.00-3.00 µM and 3.00-8.50 µM, respectively, for three amino acids. Matrix-matched six-point calibration curves showed a linear correlation coefficient (r 2) of ≥0.998. Recovery experiments validated the method by spiking the control sample at three different concentration levels (5, 50 and 100 µM), and the accuracy level was 85-110%. Except for Thr and Ser, intra- (n = 3) and inter-day (n = 3) precision fell between 0.02 and 7.28. Salivary amino acids can serve as possible biomarkers for various malignancies, with fluctuations in body fluids being crucial for cancer diagnosis; therefore, examining amino acid patterns in saliva can assist in early cancer detection. LC-MS offers improved selectivity and sensitivity for non-derivatised amino acid analysis, surpassing conventional methods and offering proactive quality assurance, making it suitable for complicated sample matrices. These discoveries could be significant in investigating new pathways and cancer treatments and looking for possible AA biomarkers for other malignancies and diseases.