Your browser doesn't support javascript.
loading
Evaluation of the DxU 850m Iris automated urine microscopy analyzer for identifying culture-negative urine samples: From a perspective of reducing urine culture number.
Stærk, Kristian; Acar, Ziyap; Gertsen, Jan Berg; Justesen, Ulrik S.
Afiliación
  • Stærk K; Department of Clinical Microbiology, Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, Odense Denmark. Electronic address: kristian.staerk@rsyd.dk.
  • Acar Z; Department of Clinical Microbiology, Odense University Hospital, Odense, Denmark.
  • Gertsen JB; Department of Clinical Microbiology, Aarhus University Hospital, Århus, Denmark.
  • Justesen US; Department of Clinical Microbiology, Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, Odense Denmark.
Diagn Microbiol Infect Dis ; 110(1): 116439, 2024 Sep.
Article en En | MEDLINE | ID: mdl-39024934
ABSTRACT
We evaluated the DxU 850m Iris Urine Microscopy analyzer as a screening tool for excluding negative urine samples (n = 1337). At a cutoff of 103 colony counts·mL-1, sensitivity was 55.1 %, specificity 68.6 %. The DxU 850m Iris does not offer acceptable prediction of culture-negative urine samples at the tested cutoff.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Orina / Sensibilidad y Especificidad / Urinálisis / Microscopía Límite: Humans Idioma: En Revista: Diagn Microbiol Infect Dis Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Orina / Sensibilidad y Especificidad / Urinálisis / Microscopía Límite: Humans Idioma: En Revista: Diagn Microbiol Infect Dis Año: 2024 Tipo del documento: Article Pais de publicación: Estados Unidos