Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices.
Feng, Yue; Nitter, Tore; Bertelsen, Geir; Stojanovic, Aleksandar.
Afiliación
  • Feng Y; Institute of Community Medicine, Faculty of Health Sciences, University in Tromsø, Tromsø, Norway.
  • Bertelsen G; Institute of Community Medicine, Faculty of Health Sciences, University in Tromsø, Tromsø, Norway.
  • Stojanovic A; Department of Ophthalmology, University Hospital North Norway, Tromsø, Norway.
Clin Exp Ophthalmol ; 2024 Jul 21.
Article en En | MEDLINE | ID: mdl-39034272
ABSTRACT

BACKGROUND:

To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices Anterion, Casia SS-1000, IOLMaster 700, and MS-39.

METHODS:

Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J0 and J45. The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within-subject standard deviation (Sw). The agreement among the four devices was assessed by pairwise comparisons and Bland-Altman plots.

RESULTS:

For all devices, the repeatability of TCA measurements showed Sw ≤0.23 D for TCA magnitude, ≤0.14 D for J0, and ≤0.12 D for J45. There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (Sw) of axis measurements had a statistically significant negative correlation with the TCA magnitude (p < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements (Sw 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 ± 0.033 mm (p < 0.001).

CONCLUSIONS:

All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.