Correction of scatter in CBCT with a new grid design.
Cui, Han; Su, Ting; Zhang, Xin; Zhu, Jiongtao; Tan, Yuhang; Zheng, Hairong; Liang, Dong; Ge, Yongshuai.
Afiliación
  • Cui H; Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
  • Su T; Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
  • Zhang X; Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
  • Zhu J; Key Laboratory of Optoelectronic Devices and Systems of Ministry of Education and Guangdong Province, College of Physics and Optoelectronic Engineering, Shenzhen University, Shenzhen 518060, People's Republic of China.
  • Tan Y; Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
  • Zheng H; Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
  • Liang D; Key Laboratory of Biomedical Imaging Science and System, Shenzhen, Shenzhen, Guangdong 518055, People's Republic of China.
  • Ge Y; Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.
Phys Med Biol ; 69(16)2024 Aug 01.
Article en En | MEDLINE | ID: mdl-39047782
ABSTRACT
Objective.This study aims at developing a simple and rapid Compton scatter correction approach for cone-beam CT (CBCT) imaging.Approach.In this work, a new Compton scatter estimation model is established based on two distinct CBCT scans one measures the full primary and scatter signals without anti-scatter grid (ASG), and the other measures a portion of primary and scatter signals with ASG. To accelerate the entire data acquisition speed, a half anti-scatter grid (h-ASG) that covers half of the full detector surface is proposed. As a result, the distribution of scattered x-ray photons could be estimated from a single CBCT scan. Physical phantom experiments are conducted to validate the performance of the newly proposed scatter correction approach.Main results.Results demonstrate that the proposed half grid approach can quickly and precisely estimate the distribution of scattered x-ray photons from only one single CBCT scan, resulting in a significant reduction of shading artifacts. In addition, it is found that the h-ASG approach is less sensitive to the grid transmission fractions, grid ratio and object size, indicating a robust performance of the new method.Significance.In the future, the Compton scatter artifacts can be quickly corrected using a half grid in CBCT imaging.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dispersión de Radiación / Procesamiento de Imagen Asistido por Computador / Fantasmas de Imagen / Tomografía Computarizada de Haz Cónico Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dispersión de Radiación / Procesamiento de Imagen Asistido por Computador / Fantasmas de Imagen / Tomografía Computarizada de Haz Cónico Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2024 Tipo del documento: Article