Differences Between Glycated Hemoglobin and Glucose Management Indicator in Real-Time and Intermittent Scanning Continuous Glucose Monitoring in Adults With Type 1 Diabetes.
J Diabetes Sci Technol
; : 19322968241262106, 2024 Jul 29.
Article
en En
| MEDLINE
| ID: mdl-39075889
ABSTRACT
BACKGROUND:
This study demonstrates the difference between glucose management indicator (GMI) and glycated hemoglobin (HbA1c) according to sensor mean glucose and HbA1c status using 2 continuous glucose monitoring (CGM) sensors in people with type 1 diabetes.METHODS:
A total of 275 subjects (117 Dexcom G6 [G6] and 158 FreeStyle Libre 1 [FL]) with type 1 diabetes was included. The G6 and FL sensors were used, respectively, over 90 days to analyze 682 and 515 glycemic profiles that coincide with HbA1c.RESULTS:
The mean HbA1c was 6.6% in Dexcom G6 and 7.2% in FL profiles. In G6 profiles, GMI was significantly higher than HbA1c irrespective of mean glucose (all P < .001, mean difference 0.58% ± 0.49%). The GMI was significantly higher than the given HbA1c when HbA1c was below 8.0% (all P < .001). The discordance was the highest at 0.9% for lower HbA1c values (5.0%-5.9%). The proportion of discordance >0.5% improved from 60.1% to 30.9% when using the revised GMI equation in G6 profiles. In FL profile, the overall mean difference between GMI and HbA1c was 0 (P = .966). The GMI was significantly lower by 0.9% than HbA1c of 9.0% to 9.9% and higher by 0.5% in HbA1c of 5.0% to 5.9% (all P < .001).CONCLUSIONS:
The GMI is overestimated in G6 users, particularly those with well-controlled diabetes, but the GMI and HbA1c discordance improved with a revised equation from the observed data. The FL profile showed greater discordance for lower HbA1c levels or higher HbA1c levels.