Biomechanical Comparison of Three Locking Compression Plate Constructs from Three Manufacturers under Cyclic Torsional Loading in a Fracture Gap Model.
Lai, Lik Hang; James, Daniel Reynolds; Appleyard, Richard Charles; Cadman, Joseph.
Afiliación
  • Lai LH; Department of Surgery, Small Animal Specialist Hospital (SASH), Sydney, New South Wales, Australia.
  • James DR; Department of Surgery, Sydney Veterinary Emergency & Specialists, Sydney, New South Wales, Australia.
  • Appleyard RC; Orthopaedic Biomechanics Research Group, Macquarie University, Sydney, New South Wales, Australia.
  • Cadman J; Orthopaedic Biomechanics Research Group, Macquarie University, Sydney, New South Wales, Australia.
Article en En | MEDLINE | ID: mdl-39102851
ABSTRACT

OBJECTIVE:

The aim of the study was to compare the stiffness and cyclic fatigue of locking compression plate constructs from three manufacturers, DePuy Synthes (DPS), Knight Benedikt (KB), and Provet Veterinary Instrumentation (Vi), under cyclic torsion.

METHODS:

The constructs of DPS, KB, and Vi were assembled by fixing a 10-hole 3.5-mm stainless steel locking compression plate 1 mm away from a validated bone model with a fracture gap of 47 mm. The corresponding drill guides and locking screws were used. Three groups of six constructs were tested in cyclic torsion until failure.

RESULTS:

There was no significant difference in initial stiffness between DPS constructs (28.83 ± 0.84 N·m/rad) and KB constructs (28.38 ± 0.81 N·m/rad), and between KB constructs and Vi constructs (27.48 ± 0.37 N·m/rad), but the DPS constructs were significantly stiffer than the Vi constructs. The DPS constructs sustained the significantly highest number of cycles (24,833 ± 2,317 cycles) compared with KB constructs (16,167 ± 1,472 cycles) and Vi constructs (19,833 ± 4,792 cycles), but the difference between KB and Vi constructs was not significant. All constructs failed by screw damage at the shaft between the plate and the bone model.

CONCLUSION:

DPS constructs showed superior initial torsional stiffness and cyclic fatigue life than Vi constructs, whereas KB and Vi constructs shared comparable results. Further investigation is required to assess the clinical significance of these biomechanical differences.