Biomechanical Effects of Multi-segment Fixation on Lumbar Spine and Sacroiliac Joints: A Finite Element Analysis.
Zhao, Geng; Wang, Lianlei; Wang, Hongwei; Li, Chao; Yuan, Suomao; Sun, Junyuan; Tian, Yonghao; Liu, Xinyu.
Afiliación
  • Zhao G; School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing, China.
  • Wang L; Department of Orthopedics, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
  • Wang H; Department of Orthopedics, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
  • Li C; Collage of Artificial Intelligence and Big Data for Medical Sciences, Shandong First Medical University, Jinan, China.
  • Yuan S; Department of Orthopedics, Peking University First Hospital, Beijing, China.
  • Sun J; Department of Orthopedics, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
  • Tian Y; Department of Orthopedics, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
  • Liu X; Department of Orthopedics, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, China.
Orthop Surg ; 16(10): 2499-2508, 2024 Oct.
Article en En | MEDLINE | ID: mdl-39118238
ABSTRACT

OBJECTIVE:

Spine fixation surgery affects the biomechanical environment in the sacroiliac joint (SIJ), which may lead to the SIJ pain or degeneration after surgery. The purpose of this study is to determine the impact of the number and position of fixed segments on the SIJs and provide references for surgeons to plan fixation levels and enhance surgical outcomes.

METHODS:

The intact lumbar-pelvis finite element (FE) models and 11 fixation FE models with different number and position of fixed segments were developed based on CT images. A 400N follower load and 10° range of motion (ROM) of the spine were applied to the superior endplate of L1 to simulate the flexion, extension, bending and torsion motion after surgery. The peak stress on the SIJs, lumbar intervertebral discs, screws and rods were calculated to evaluate the biomechanical effects of fixation procedures.

RESULTS:

With the lowermost instrumented vertebra (LIV) of L5 or S1, the peak stress on SIJs increased with the number of fixed segments increasing. The flexion motion led to the greater von Mises stress on SIJ compared with other load conditions. Compared with the intact model, peak stress on all fixed intervertebral discs was reduced in the models with less than three fixed segments, and it increased in the models with more than three fixed segments. The stress on the SIJ was extremely high in the models with all segments from L1 to L5 fixed, including L1-L5, L1-S1 and L1-S2 fixation models. The stress on the segment adjacent to the fixed segments was significant higher compared to that in the intact model. The peak stress on rods and screws also increased with the number of fixed segments increasing in the flexion, extension and bending motion, and the bending and flexion motions led to the greater von Mises stress on SIJs.

CONCLUSION:

Short-term fixation (≤2 segments) did not increase the stress on the SIJs significantly, while long-term segment fixation (≥4 segments) led to greater stress on the SIJs especially when all the L1-L5 segments were fixed. Unfixed lumbar segments compensated the ROM loss of the fixed segments, and the preservation of lumbar spine mobility would reduce the risks of SIJ degeneration.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Articulación Sacroiliaca / Fusión Vertebral / Rango del Movimiento Articular / Análisis de Elementos Finitos / Vértebras Lumbares Límite: Adult / Humans / Male Idioma: En Revista: Orthop Surg Año: 2024 Tipo del documento: Article País de afiliación: China Pais de publicación: Australia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Articulación Sacroiliaca / Fusión Vertebral / Rango del Movimiento Articular / Análisis de Elementos Finitos / Vértebras Lumbares Límite: Adult / Humans / Male Idioma: En Revista: Orthop Surg Año: 2024 Tipo del documento: Article País de afiliación: China Pais de publicación: Australia