Your browser doesn't support javascript.
loading
A method to determine the retinal radiant exposure caused by ophthalmoscopy.
Campbell, Charles.
Afiliación
  • Campbell C; Berkeley, California, USA.
Article en En | MEDLINE | ID: mdl-39127914
ABSTRACT

PURPOSE:

To introduce a method to calculate retinal irradiance caused by ophthalmoscopy. This may be used to verify the compliance of an instrument with the radiation limits set by light hazard standards. The proposed method is simpler to use and less prone to error than the methods currently found in the light hazard standards.

METHODS:

The optical properties of the standardised human eye, specified by current light hazard standards, are used to calculate the magnification of an aerial image of the retinal surface by the combination of the optics of eye and the auxiliary lens used for ophthalmoscopy. The magnification of the aerial image is used to transform the spectral irradiance of the instrument illumination source to retinal irradiation values. The spectral irradiance of the instrument illumination source can be measured directly as the aerial image is located in the focal plane of the viewing optics of the ophthalmoscope. These spectral irradiation values are then processed using weightings specified by current light hazard standards to give a weighted irradiance which is converted directly to a retinal irradiance value.

RESULTS:

A single formula is provided to calculate the retinal irradiance using the processed, measured spectral irradiance values of the illumination source.

CONCLUSION:

The new method introduced here is simpler to use, requires fewer physical measurements and is less likely to introduce measurement error than that currently found in light hazard standards. The only physical measurement that needs to be taken is the illumination source spectral irradiance measured in the viewing focal plane of the instrument. These values are weighted using given in the light hazard standards. The combined irradiance value is then converted to retinal irradiance using the formula given in this paper.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Ophthalmic Physiol Opt Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Reino Unido