Interpretive standards and quality control guidelines for cefpiramide disk susceptibility tests.
J Clin Microbiol ; 22(4): 688-90, 1985 Oct.
Article en En | MEDLINE | ID: mdl-4077974
Disk susceptibility tests with 30- and 75-micrograms cefpiramide disks were evaluated with 614 bacterial isolates. Quality control parameters were also evaluated, and control limits for disk tests are recommended. Tests with 75-micrograms disks are recommended, with zone size standards of greater than or equal to 19 mm for susceptible (MIC, less than or equal to 32 micrograms/ml) and less than or equal to 15 mm for resistant (MIC, greater than or equal to 128 micrograms/ml).
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Pruebas de Sensibilidad Microbiana / Cefalosporinas Tipo de estudio: Guideline Idioma: En Revista: J Clin Microbiol Año: 1985 Tipo del documento: Article Pais de publicación: Estados Unidos