Your browser doesn't support javascript.
loading
Radiation dose in neurological computed tomographic scanning.
Radiol Technol ; 51(1): 21-6, 1979.
Article en En | MEDLINE | ID: mdl-472223
ABSTRACT
Patient dose and dose distribution during neurological computed tomography examinations were determined with five different computed tomography scanners. Maximum intracranial doses ranged from 1.17 to 2.67 rads. Doses to the lens of the eye ranged from 0.23 to 2.81 rads. These levels are considered and compared with patient doses reported for other computed tomography studies and for conventional tomographic examinations. In general, patient dose during computer tomographic examinations is less than one quarter of that during conventional tomography of the head.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Encéfalo / Tomografía Computarizada por Rayos X Límite: Humans Idioma: En Revista: Radiol Technol Año: 1979 Tipo del documento: Article
Buscar en Google