A precision isotonic measuring system for isolated tissues.
J Pharmacol Methods
; 12(4): 259-64, 1984 Dec.
Article
en En
| MEDLINE
| ID: mdl-6536830
An isotonic measuring system is described which utilizes an angular position transducer of the linear differential voltage transformer type. Resistance to corrosion, protection against the ingress of solutions, and ease of mounting and setting up were the mechanical objectives. Accuracy, linearity, and freedom from drift were essential requirements of the electrical specification. A special housing was designed to accommodate the transducer to overcome these problems. A control unit incorporating a power supply and electronic filtering components was made to serve up to four such transducers. The transducer output voltage is sufficiently high to drive directly even low sensitivity chart recorders. Constructional details and a circuit diagram are included. Fifty such transducers have been in use for up to four years in these laboratories. Examples of some of the published work done using this transducer system are referenced.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Contracción Isotónica
/
Contracción Muscular
Límite:
Animals
Idioma:
En
Revista:
J Pharmacol Methods
Año:
1984
Tipo del documento:
Article
Pais de publicación:
Estados Unidos