Evaluation of the TRI 'dipstick' test for the detection of drugs of abuse in urine.
J Anal Toxicol
; 5(1): 14-9, 1981.
Article
en En
| MEDLINE
| ID: mdl-6783782
An evaluation of Technology Resources Inc. (TRI) Amphetamine, Barbiturate, Narcotic (G) and Narcotic (S) "Dipsticks" for drugs of abuse in urine was made. The results obtained by six individuals reading the "Dipstick" papers was compared with the analysis of the same urine samples, by a combination of TLC, EMIT, RIA and GLC. The data obtained with "Dipstick" papers, regardless of the drug tested, were clearly unreliable (high percentage of false negatives, low percentage of true positives) and the assay was unsuitable as a technique for screening urines for drugs of abuse.