Penetration of the distal femur by the distal locking device of Brooker Wills interlocking nail.
Ebraheim, N A; Paley, K J.
Afiliación
  • Ebraheim NA; Department of Orthopaedic Surgery, Medical College of Ohio, Toledo 43699-0008.
Clin Orthop Relat Res ; (297): 218-23, 1993 Dec.
Article en En | MEDLINE | ID: mdl-8242934
ABSTRACT
Of seven patients with complications caused by penetration of the distal locking device through the distal femur, three had penetration of the distal locking device into the knee joint. Two patients had penetration into the patellofemoral joint. One patient had penetration into the lateral cortex of the distal femur and another patient had penetration into the medial cortex of the distal femur. All but one patient required removal of the rods and/or the distal locking device. The main causes of complications include rod rotation, previous surgery, preexisting deformity of the femur, and failure to insert the rod into its proper location distally.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Clavos Ortopédicos / Fracturas del Fémur / Fijación Intramedular de Fracturas Tipo de estudio: Diagnostic_studies / Etiology_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Año: 1993 Tipo del documento: Article
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Clavos Ortopédicos / Fracturas del Fémur / Fijación Intramedular de Fracturas Tipo de estudio: Diagnostic_studies / Etiology_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Orthop Relat Res Año: 1993 Tipo del documento: Article