Sensitive high-performance liquid chromatographic fluorescence assay for the quantitation of topotecan (SKF 104864-A) and its lactone ring-opened product (hydroxy acid) in human plasma and urine.
Loos, W J; Stoter, G; Verweij, J; Schellens, J H.
Afiliación
  • Loos WJ; Department of Medical Oncology, Rotterdam Cancer Institute/University Hospital Rotterdam, Netherlands.
J Chromatogr B Biomed Appl ; 678(2): 309-15, 1996 Apr 12.
Article en En | MEDLINE | ID: mdl-8738036
A sensitive reversed-phase high-performance liquid chromatographic fluorescence method is described for the simultaneous determination of topotecan (I) and the hydrolysed lactone ring-opened product hydroxy acid (II) in plasma and for the determination of I in urine. To 250 microliters of plasma, a 750-microliters volume of cold methanol was added to stabilize the pH-dependent conversion of I into II. In plasma, the lower limit of quantitation (LLQ) for both compounds was 0.10 ng/ml. The between-day variation for I at the LLQ was 7.1% and for II was 5.5%. Prior to injection, urine samples were acidified with orthophosphoric acid and diluted with phosphate-buffered saline (PBS). In urine, the calibration curve for I was linear in the range of 10 to 250 ng/ml and the LLQ was 10 ng/ml. The assay was developed to enable pharmacological analysis of I, in on-going phase I and II studies, in patients with solid tumors.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Camptotecina / Cromatografía Líquida de Alta Presión / Inhibidores Enzimáticos Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Chromatogr B Biomed Appl Año: 1996 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: Países Bajos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Camptotecina / Cromatografía Líquida de Alta Presión / Inhibidores Enzimáticos Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: J Chromatogr B Biomed Appl Año: 1996 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: Países Bajos