Comparison of arterial wall reaction after passage of the Hydrolyser device versus a thrombectomy balloon in an animal model.
van Ommen, V G; van der Veen, F H; Geskes, G G; Daemen, M; Habets, J; Dassen, W R; Wellens, H J.
Afiliación
  • van Ommen VG; Department of Cardiology, University Hospital of Maastricht, The Netherlands.
J Vasc Interv Radiol ; 7(3): 451-4, 1996.
Article en En | MEDLINE | ID: mdl-8761831
ABSTRACT

PURPOSE:

This study was designed to compare the reaction of the vessel wall after application of the Hydrolyser hydrodynamic thrombectomy device to the reaction after use of a balloon thrombectomy catheter. The influence of the vessel inner diameter on vessel wall reaction was evaluated after passage of the Hydrolyser. MATERIALS AND

METHODS:

After measurement of the vessel inner diameter with intravascular ultrasound (US), 102 segments of femoral and carotid arteries of goats were treated with one of the following four procedures passage of the intravascular US catheter alone; passage of the Hydrolyser without or with an activated jet; or passage of an inflated thrombectomy balloon. Histologic evaluation was performed after 3 weeks.

RESULTS:

Intimal thickening (more than five cell layers of neointima) 3 weeks after treatment occurred more frequently after passage of the balloon than after any of the other procedures (P < .001). For vessels with a diameter of 3-4 mm, 4-5mm, or more than 5 mm, no significant difference in vessel wall reaction was observed following Hydrolyser passage.

CONCLUSION:

In this model, passage of the Hydrolyser device resulted in less intimal reaction compared with the thrombectomy balloon.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Endotelio Vascular / Cateterismo / Reacción a Cuerpo Extraño / Trombectomía Límite: Animals Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 1996 Tipo del documento: Article País de afiliación: Países Bajos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Endotelio Vascular / Cateterismo / Reacción a Cuerpo Extraño / Trombectomía Límite: Animals Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 1996 Tipo del documento: Article País de afiliación: Países Bajos
...