Effect of different surface finishing and of hydroxyapatite coatings on passive and corrosion current of Ti6Al4V alloy in simulated physiological solution.
Biomaterials
; 18(11): 783-7, 1997 Jun.
Article
en En
| MEDLINE
| ID: mdl-9177856
Direct and alternating current electrochemical tests were carried out on Ti6Al4V with different surface finishing and with hydroxyapatite (HA) coatings. Sand-blasting and rough titanium deposits obtained by vacuum plasma spraying (VPS) bring about an increase of passive and corrosion current density (c.d.) with respect to smooth Ti6Al4V, as a consequence of the augmentation of the real surface. The presence of HA deposits obtained by VPS causes an increase of passive and corrosion c.d. of the metallic substrate of about one order of magnitude and this should be taken into account in view of human body applications.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Titanio
/
Aleaciones
/
Hidroxiapatitas
Tipo de estudio:
Prognostic_studies
Límite:
Humans
Idioma:
En
Revista:
Biomaterials
Año:
1997
Tipo del documento:
Article
País de afiliación:
Italia
Pais de publicación:
Países Bajos