Nonuniformity intrinsic flood artifact caused by point-source syringe needle.
Hackett, M T; Magoun, S L; Thompson, R B.
Afiliación
  • Hackett MT; Nuclear Medicine Service, Department of Veterans Affairs Medical Centers, Lexington, Kentucky 40511-1093, USA.
J Nucl Med Technol ; 25(1): 41-3, 1997 Mar.
Article en En | MEDLINE | ID: mdl-9239602
ABSTRACT
UNLABELLED The purpose of this study was to investigate the cause of a nonuniformity artifact seen on an intrinsic flood.

METHODS:

Intrinsic floods were acquired on three different cameras using a point source in a 1-cc syringe, with and without a needle attached. Floods were obtained with the syringe placed perpendicular and parallel to the camera face.

RESULTS:

A nonuniformity artifact was evident if a syringe with a needle was placed perpendicular to the crystal face.

CONCLUSION:

If a syringe is used as a point source during the acquisition of an intrinsic quality control flood, the needle should not be placed perpendicular to the camera face.