Nonuniformity intrinsic flood artifact caused by point-source syringe needle.
J Nucl Med Technol
; 25(1): 41-3, 1997 Mar.
Article
en En
| MEDLINE
| ID: mdl-9239602
ABSTRACT
UNLABELLED The purpose of this study was to investigate the cause of a nonuniformity artifact seen on an intrinsic flood. METHODS:
Intrinsic floods were acquired on three different cameras using a point source in a 1-cc syringe, with and without a needle attached. Floods were obtained with the syringe placed perpendicular and parallel to the camera face.RESULTS:
A nonuniformity artifact was evident if a syringe with a needle was placed perpendicular to the crystal face.CONCLUSION:
If a syringe is used as a point source during the acquisition of an intrinsic quality control flood, the needle should not be placed perpendicular to the camera face.