Your browser doesn't support javascript.
loading
Power injection of peripherally inserted central catheters.
Rivitz, S M; Drucker, E A.
Afiliación
  • Rivitz SM; Department of Radiology, Massachusetts General Hospital, Boston, USA.
J Vasc Interv Radiol ; 8(5): 857-63, 1997.
Article en En | MEDLINE | ID: mdl-9314379
PURPOSE: To study the tolerance of peripherally inserted central catheters (PICCs) of varying sizes and materials to power injection of radiographic contrast agents. MATERIALS AND METHODS: Eight different models of silicone and five different models of polyurethane single-lumen PICCs were injected with increasing rates of iothalamate 60% with use of a power injector. Tolerated and bursting rates and pressures were recorded. RESULTS: There was a wide range of tolerated rates and pressures, depending on the inner and outer diameters of the catheters and on the catheter material. Silicone PICCs tolerated rates between 0.4 and 7.0 mL/sec and polyurethane PICCs tolerated rates between 0.6 and 10.2 mL/sec, depending on the specific catheter. The 5-F silicone PICCs and the 4-F and 5-F polyurethane PICCs tested all tolerated rates greater than 4 mL/sec. Silicone catheters tolerated pressures between 107 and 184 psi, and polyurethane catheters tolerated pressures between 160 and 314 psi. CONCLUSIONS: Larger single-lumen silicone and polyurethane PICCs may be suitable for power injection of contrast agents.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cateterismo Venoso Central / Medios de Contraste / Ácido Yotalámico Límite: Humans Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cateterismo Venoso Central / Medios de Contraste / Ácido Yotalámico Límite: Humans Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos