Your browser doesn't support javascript.
loading
Simultaneous PET and MR imaging.
Shao, Y; Cherry, S R; Farahani, K; Meadors, K; Siegel, S; Silverman, R W; Marsden, P K.
Afiliación
  • Shao Y; Crump Institute for Biological Imaging, Department of Molecular and Medical Pharmacology, UCLA School of Medicine 90095, USA.
Phys Med Biol ; 42(10): 1965-70, 1997 Oct.
Article en En | MEDLINE | ID: mdl-9364592
ABSTRACT
We have developed a prototype PET detector which is compatible with a clinical MRI system to provide simultaneous PET and MR imaging. This single-slice PET system consists of 48 2 x 2 x 10 mm3 LSO crystals in a 38 mm diameter ring configuration that can be placed inside the receiver coil of the MRI system, coupled to three multi-channel photomultipliers housed outside the main magnetic field via 4 m long and 2 mm diameter optical fibres. The PET system exhibits 2 mm spatial resolution, 41% energy resolution at 511 keV and 20 ns timing resolution. Simultaneous PET and MR phantom images were successfully acquired.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Tomografía Computarizada de Emisión / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies Idioma: En Revista: Phys Med Biol Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Tomografía Computarizada de Emisión / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies Idioma: En Revista: Phys Med Biol Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos