Assessment of the CAMINO intracranial pressure device in clinical practice.
Schürer, L; Münch, E; Piepgras, A; Weigel, R; Schilling, L; Schmiedek, P.
Afiliación
  • Schürer L; Department of Neurosurgery, Klinikum Mannheim, University of Heidelberg, Federal Republic of Germany.
Acta Neurochir Suppl ; 70: 296-8, 1997.
Article en En | MEDLINE | ID: mdl-9416352
The purpose of this study was to investigate reliability, handling characteristics and complication rate of the CAMINO-ICP-monitor-system in clinical routine. In a case controlled study 82 patients with intracranial pathology necessitating ICP-monitoring received either a ventricular or a parenchymal CAMINO-device. Clinical assessment of curve shape and apparent reliability of the measurement was documented. Probe position and presence of hematoma was evaluated in all patients with a CT after probe insertion. Handling complications, i.e. dislocation were recorded. At the end of the measuring period the drift of the probe was checked ex vivo and a two point calibration was performed using a water column. During one year 82 patients received 95 probes (parench, 73. ventric. 22). The average measuring period was 91.3 +/- 70.6 hrs. Catheter position was verified by CCT for 67 (70.5%) probes. 92.5% of the devices were placed correctly. Clinically 88.4% of the measurements were assessed plausible, in 8.2% the displayed ICP-values were judged to be too high, in 2.1% too low. Probe drift after explanation was -0.21 mmHg/24 hrs. The mean value of the recalibrated probes in the water column corresponding to 15.8 mmHg was 14.7 +/- 1.9 mmHg. There was no correlation between neither drift nor function in the water column and the duration of the measurement. Technical complications exclusively related to the construction of the CAMINO-system like kinking of the cable, dislocation (probe pulled out) or dislocated fixation screw were too high (25.3%).
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Presión Intracraneal / Tecnología de Fibra Óptica / Monitoreo Fisiológico Tipo de estudio: Diagnostic_studies / Evaluation_studies / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Child / Humans / Middle aged Idioma: En Revista: Acta Neurochir Suppl Año: 1997 Tipo del documento: Article Pais de publicación: Austria
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Presión Intracraneal / Tecnología de Fibra Óptica / Monitoreo Fisiológico Tipo de estudio: Diagnostic_studies / Evaluation_studies / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Child / Humans / Middle aged Idioma: En Revista: Acta Neurochir Suppl Año: 1997 Tipo del documento: Article Pais de publicación: Austria