Liver failure and death after exposure to microcystins at a hemodialysis center in Brazil.
Jochimsen, E M; Carmichael, W W; An, J S; Cardo, D M; Cookson, S T; Holmes, C E; Antunes, M B; de Melo Filho, D A; Lyra, T M; Barreto, V S; Azevedo, S M; Jarvis, W R.
Afiliación
  • Jochimsen EM; Hospital Infections Program, National Center for Infectious Diseases, Centers for Disease Control and Prevention, Atlanta, GA 30333, USA.
N Engl J Med ; 338(13): 873-8, 1998 Mar 26.
Article en En | MEDLINE | ID: mdl-9516222
ABSTRACT

BACKGROUND:

Hemodialysis is a common but potentially hazardous procedure. From February 17 to 20, 1996, 116 of 130 patients (89 percent) at a dialysis center (dialysis center A) in Caruaru, Brazil, had visual disturbances, nausea, and vomiting associated with hemodialysis. By March 24, 26 of the patients had died of acute liver failure.

METHODS:

A case patient was defined as any patient undergoing dialysis at dialysis center A or Caruaru's other dialysis center (dialysis center B) during February 1996 who had acute liver failure. To determine the risk factors for and the source of the outbreak, we conducted a cohort study of the 130 patients at dialysis center A and the 47 patients at dialysis center B, reviewed the centers' water supplies, and collected water, patients' serum, and postmortem liver tissue for microcystin assays.

RESULTS:

One hundred one patients (all at dialysis center A) met the case definition, and 50 died. Affected patients who died were older than those who survived (median age, 47 vs. 35 years, P<0.001). Furthermore, all 17 patients undergoing dialysis on the Tuesday-, Thursday-, and Saturday-night schedule became ill, and 13 of them (76 percent) died. Both centers received water from a nearby reservoir. However, the water supplied to dialysis center B was treated, filtered, and chlorinated, whereas the water supplied to dialysis center A was not. Microcystins produced by cyanobacteria were detected in water from the reservoir and from dialysis center A and in serum and liver tissue of case patients.

CONCLUSIONS:

Water used for hemodialysis can contain toxic materials, and its quality should therefore be carefully monitored.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Péptidos Cíclicos / Toxinas Bacterianas / Contaminantes Químicos del Agua / Abastecimiento de Agua / Diálisis Renal / Fallo Hepático Agudo Tipo de estudio: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Límite: Adult / Humans / Middle aged País/Región como asunto: America do sul / Brasil Idioma: En Revista: N Engl J Med Año: 1998 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Péptidos Cíclicos / Toxinas Bacterianas / Contaminantes Químicos del Agua / Abastecimiento de Agua / Diálisis Renal / Fallo Hepático Agudo Tipo de estudio: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Límite: Adult / Humans / Middle aged País/Región como asunto: America do sul / Brasil Idioma: En Revista: N Engl J Med Año: 1998 Tipo del documento: Article País de afiliación: Estados Unidos