Your browser doesn't support javascript.
loading
Is atrial sensing of ventricular far-field signals important in single-lead VDD pacing?
Nowak, B; Kramm, B; Schwaier, H; Maimaitiming, A; Geil, S; Zellerhoff, C; Przibille, O; Rosocha, S; Liebrich, A; Meyer, J.
Afiliación
  • Nowak B; Medical Clinic, Johannes Gutenberg University, Mainz, Germany. Nowak@2-med.klinik.unimainz.de
Pacing Clin Electrophysiol ; 21(11 Pt 2): 2236-9, 1998 Nov.
Article en En | MEDLINE | ID: mdl-9825325
ABSTRACT
UNLABELLED In single-lead VDD pacing the atrial sensitivity frequently is programmed to sensitive values. Atrial sensing of ventricular far-field signals should be reduced by differential atrial sensing. The aim of the study was to evaluate the effectiveness of this approach.

METHODS:

The study included 10 patients with a single-lead VDD pacemaker (Thera 8948, Lead 5032). The atrial sensitivity was set to its most sensitive value of 0.18 mV and the telemetered intraatrial EGM was continuously recorded. After atrial tracked ventricular pacing, VVI pacing was performed with pacing rates from 100 to 160 beats/min in steps of 10 beats/min and up to 165 beats/min. The peak-to-peak amplitudes of P waves (P) and ventricular far-field signals (VFFS) were measured from the recordings. The ratio P/VFFS that defines the atrial signal-to-noise ratio was calculated, and the time from stimulus to maximum of the far-field signals amplitude (Tmax) was measured.

RESULTS:

P measured 0.98 +/- 0.76 mV. A VFFS was visible in the atrial channel in all patients with an amplitude of 0.45 +/- 0.25 mV (range 0.01-1.0 mV), independent of the pacing rate. The ratio P/VFFS was 3.9 +/- 4.2 (range 0.9-21.0). Tmax measured 99.4 +/- 15.2 ms during sinus rhythm. A rate dependent shortening of Tmax to 92.7 +/- 11.2 ms at 140 beats/min was observed (P = 0.001). At rates above 140 beats/min no further shortening occurred.

CONCLUSION:

Ventricular far-field signals are measurable in the atrial channel of VDD systems and may reach considerable amplitudes, which are not rate dependent. Although differential sensing provides favorable P waves to ventricular far-field signal ratios, refractory periods are needed to avoid far-field sensing. The rate dependent shortening of the ventricular signal can be detected in the atrial channel in VDD pacing.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Bloqueo Cardíaco Límite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Año: 1998 Tipo del documento: Article País de afiliación: Alemania
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Bloqueo Cardíaco Límite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Año: 1998 Tipo del documento: Article País de afiliación: Alemania