Performance characteristics of a Kodak computed radiography system.
Bradford, C D; Peppler, W W; Dobbins, J T.
Afiliación
  • Bradford CD; Department of Medical Physics, University of Wisconsin, Madison 53706, USA.
Med Phys ; 26(1): 27-37, 1999 Jan.
Article en En | MEDLINE | ID: mdl-9949395
ABSTRACT
The performance characteristics of a photostimulable phosphor based computed radiographic (CR) system were studied. The modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE) of the Kodak Digital Science computed radiography (CR) system (Eastman Kodak Co.-model 400) were measured and compared to previously published results of a Fuji based CR system (Philips Medical Systems-PCR model 7000). To maximize comparability, the same measurement techniques and analysis methods were used. The DQE at four exposure levels (30, 3, 0.3, 0.03 mR) and two plate types (standard and high resolution) were calculated from the NPS and MTF measurements. The NPS was determined from two-dimensional Fourier analysis of uniformly exposed plates. The presampling MTF was determined from the Fourier transform (FT) of the system's finely sampled line spread function (LSF) as produced by a narrow slit. A comparison of the slit type ("beveled edge" versus "straight edge") and its effect on the resulting MTF measurements was also performed. The results show that both systems are comparable in resolution performance. The noise power studies indicated a higher level of noise for the Kodak images (approximately 20% at the low exposure levels and 40%-70% at higher exposure levels). Within the clinically relevant exposure range (0.3-3 mR), the resulting DQE for the Kodak plates ranged between 20%-50% lower than for the corresponding Fuji plates. Measurements of the presampling MTF with the two slit types have shown that a correction factor can be applied to compensate for transmission through the relief edges.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Interpretación de Imagen Radiográfica Asistida por Computador / Intensificación de Imagen Radiográfica Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: Med Phys Año: 1999 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Interpretación de Imagen Radiográfica Asistida por Computador / Intensificación de Imagen Radiográfica Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: Med Phys Año: 1999 Tipo del documento: Article País de afiliación: Estados Unidos