Comparison of FP-7 and S-2 Ahmed glaucoma valve implantation in refractory glaucoma patients for short-term follow-up / 中华医学杂志(英文版)
Chinese Medical Journal ; (24): 1128-1133, 2011.
Article en En | WPRIM | ID: wpr-239879
Biblioteca responsable: WPRO
ABSTRACT
<p><b>BACKGROUND</b>Ahmed glaucoma valves (AGV) has been used for decades, but there is no detailed report about the efficacy of AGV in Chinese glaucoma patients. This study aimed to compare the intraocular pressure (IOP) lowering efficacy and side effects of S-2 polypropylene and PF-7 silicone AGV implantation in Chinese refractory glaucoma patients.</p><p><b>METHODS</b>Patients were divided into S-2 model AGV group and FP-7 model AGV group. The complete and qualified surgical success rate, change of IOP, number of anti-glaucoma medications used and postoperative complications were recorded and analyzed.</p><p><b>RESULTS</b>Average follow-up time was comparable between two groups. IOP was reduced from (37.9 ± 12.7) mmHg preoperatively to (17.3 ± 5.3) mmHg at the last follow-up in S-2 group and reduced from (39.9 ± 14.4) mmHg to (17.7 ± 4.9) mmHg in FP-7 group. Anti-glaucoma medications were reduced from 3.8 ± 0.2 to 1.5 ± 0.2 in S-2 group, and 3.5 ± 0.2 to 0.7 ± 0.2 in FP-7 groups. The cumulative success rates were comparable in two groups, which were 61.2% and 72.1% in S-2 group and FP-7 group respectively. When IOP reduction criteria was used, complete success rates were 30.6% and 51.2% for S-2 and FP-7 groups, and qualified success rates were 86.1% and 92.7% separately. In both groups, the major complication was hypotony, and the previous trabeculectomy of patients was the major risk factor for surgery failure.</p><p><b>CONCLUSIONS</b>In this short-term retrospective study, S-2 AGV is showed at least as effective as FP-7 AGV in IOP reduction, but associated with higher rate of complications. Previous trabeculectomy is a principle risk factor for AGV implantation failure. These clinical outcomes are important for converting use of the FP-7 silicon AGV in Chinese refractory glaucoma patients.</p>
Asunto(s)
Texto completo: 1 Base de datos: WPRIM Asunto principal: Fisiología / Terapéutica / Glaucoma / Estudios Retrospectivos / Implantes de Drenaje de Glaucoma / Presión Intraocular Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Chinese Medical Journal Año: 2011 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Asunto principal: Fisiología / Terapéutica / Glaucoma / Estudios Retrospectivos / Implantes de Drenaje de Glaucoma / Presión Intraocular Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Chinese Medical Journal Año: 2011 Tipo del documento: Article