Study on the design and detection method of a novel blood pressure measurement system / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 173-177, 2013.
Article
en Zh
| WPRIM
| ID: wpr-264242
Biblioteca responsable:
WPRO
ABSTRACT
A novel blood pressure measurement system was designed which based on auscultatory method. And the electret sensor that embedded into the internal instrument can detect the Korotkoff-sound signal directly which is coupled by the cuff and transmitted in the cross connection. The BP values identification algorithm is based on combined detection of Korotkoff-sound and pulse signal, and the products of amplitudes and energies are calculated as the characteristic values of Korotkoff-sound, and the Korotkoff-sound phases are classified and detected by means of clustering of characteristic values, and then BP parameters are determined. The contrast test and statistical analysis showed good consistency and accuracy between the new BP detection method and conventional mercury sphygmomanometer.
Texto completo:
1
Base de datos:
WPRIM
Asunto principal:
Presión Sanguínea
/
Determinación de la Presión Sanguínea
/
Diseño de Equipo
/
Métodos
Tipo de estudio:
Diagnostic_studies
/
Prognostic_studies
Idioma:
Zh
Revista:
Chinese Journal of Medical Instrumentation
Año:
2013
Tipo del documento:
Article