Animal experimental research of the endothelialization of home-made atrial septal defect occluder device / 介入放射学杂志
Journal of Interventional Radiology
; (12)2001.
Article
en Zh
| WPRIM
| ID: wpr-571521
Biblioteca responsable:
WPRO
ABSTRACT
Objective To evaluate the endothelialization of Chinese nitinol atrial septal defect occlud-er device. Methods Atrial septal defect with controllable size was created by the Brockenborough needle and Rashkind balloon atrial septostomy, the occluder devices were implanted in six piglets (mean weight 7. 5 kg). Two pigs were killed each time after 1 month, 3 months and 6 months after the device implantation and then the explanted devices were examined by scanning electron microscope (SEM). Results The devices were found covering with collagen fibrosis together with diffuse endothelial cells spreading over the primer 1 month after implantation. The implants were covered mostly by ncointima 3 months after implantation and completely covered by confluent endothelial cells 6 months after the implantation. Endothelial cells were not found on the smooth marker band at 3 months, however, did exist by 6 months. Conclusions Home-made atriatseptal defect occluder devices were mostly endothelialised 3 months after the implantation and did completely at 6months.
Texto completo:
1
Base de datos:
WPRIM
Idioma:
Zh
Revista:
Journal of Interventional Radiology
Año:
2001
Tipo del documento:
Article