Shoulder hemiarthroplasty assisted by individualized navigation templates 3D printed / 中华创伤骨科杂志
Chinese Journal of Orthopaedic Trauma
; (12): 596-602, 2017.
Article
en Zh
| WPRIM
| ID: wpr-612064
Biblioteca responsable:
WPRO
ABSTRACT
Objective To investigate the efficacy and accuracy of 3D printing individualized navigation templates used in shoulder hemi-arthroplasty (HA).Methods We collected 9 adult cadaveric specimens of 18 shoulders with no obvious deformities or defects.After CT scanning was conducted on bilateral shoulders,the raw CT data in DICOM format were imported into software Mimics 10.01 and Imageware 12.0 to reconstruct 3D shoulder models.The retroversion angles and heights of bilateral humeral heads were measured to design individualized navigational templates using software Geomagic Studio 12.0.Rapid prototyping (RP) technique was used to manufacture the individualized navigation templates which were then used to assist shoulder HA in cadaveric specimens.CT scanning was performed postoperatively to measure the retroversion angle and height of the prosthesis implanted for comparison with the preoperative measurements.Results Before shoulder replacement in the 9 cadaveric specimens,respectively,the heights of left and right humeral heads averaged 300.0 ± 11.6 mm and 300.0 ± 11.4 mm,and the retroversion angles 15.6°±4.9° and 15.9°± 4.9°.After the replacement,the heights of left and right humeral heads averaged 300.0 ± 11.3 mm and 300.0 ± 11.7 mm,respectively,and the retroversion angles 15.7° ± 4.9° and 15.8° ± 5.2°,respectively.There were no significant differences between the left and right sides in the pre-or post-operative heights or retroversion angles of the humeral head (P > 0.05).There were no significant differences either between preand post-operation in the heights or retroversion angles of the left and right humeral heads (P > 0.05).The mean errors for the preoperative and postoperative heights and retroversion angles of the humeral heads were 1.44 ± 0.96 mm and 0.25° ± 0.15°,respectively.Conclusion Individualized navigation templates 3D printed can be used to assist effective and accurate shoulder HA because they are designed and manufactured according to accurate measurements of retroversion angles and heights of the humeral heads which are acquired by computer preoperatively.
Texto completo:
1
Base de datos:
WPRIM
Idioma:
Zh
Revista:
Chinese Journal of Orthopaedic Trauma
Año:
2017
Tipo del documento:
Article