ABSTRACT
BACKGROUND AND

OBJECTIVES:

Wrist-worn wearable devices provide heart rate (HR) monitoring function via photoplethysmography technology. Recently, these devices have been used by patients to measure the HR when palpitation occurs, but few validation studies of these instruments have been conducted. We assessed the accuracy of these devices for measuring a HR.

METHODS:

This study enrolled 51 consecutive patients with a history of paroxysmal supraventricular tachyarrhythmia (SVT) or paroxysmal palpitations who were scheduled to undergo an electrophysiological study (EPS). Three devices were assessed Apple Watch Series 2 (Apple), Samsung Galaxy Gear S3 (Galaxy), and Fitbit Charge 2 (Fitbit). Patients were randomly assigned to wear 2 different devices. The HR at baseline and induced SVT were measured during the EPS. After successful ablation of SVT, HR measurements was also done during atrial and ventricular pacing study.

RESULTS:

The mean patient age was 44.4±16.6 years and 27 patients were male (53%). The accuracy (within ±5 beats per minute [bpm] of an electrocardiogram [ECG] measurement) of the baseline HR measurements was 100%, 100%, and 94%, for Apple, Galaxy, and Fitbit, respectively. The HR during induced SVT ranged from 108 bpm to 228 bpm and the accuracy (within ±10 bpm of an ECG) was 100%, 90%, and 87% for the Apple, Galaxy, and Fitbit, respectively. During pacing study, accuracy of these devices was also acceptable but tended to decrease as the HR increased, and showed differences between the devices.

CONCLUSIONS:

Wrist-worn wearable devices accurately measure baseline and induced SVT HR. TRIAL REGISTRATION Clinical Research Information Service Identifier KCT0002282
Asunto(s)

Texto completo: Disponible Base de datos: WPRIM (Pacífico Occidental) Asunto principal: Taquicardia / Taquicardia Supraventricular / Fotopletismografía / Electrocardiografía / Galaxias / Corazón / Frecuencia Cardíaca / Servicios de Información Tipo de estudio: Estudio pronóstico Límite: Humanos / Masculino Idioma: Inglés Revista: Korean Circulation Journal Año: 2019 Tipo del documento: Artículo
Texto completo: Disponible Base de datos: WPRIM (Pacífico Occidental) Asunto principal: Taquicardia / Taquicardia Supraventricular / Fotopletismografía / Electrocardiografía / Galaxias / Corazón / Frecuencia Cardíaca / Servicios de Información Tipo de estudio: Estudio pronóstico Límite: Humanos / Masculino Idioma: Inglés Revista: Korean Circulation Journal Año: 2019 Tipo del documento: Artículo
...