ABSTRACT
OBJECTIVE@#In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.@*METHODS@#A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.@*RESULTS@#The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.@*CONCLUSIONS@#To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.
Texto completo:
Disponible
Contexto en salud:
Agenda de Salud Sostenible para las Américas
/
ODS3 - Salud y Bienestar
Problema de salud:
Objetivo 9: Enfermedades no transmisibles y salud mental
/
Meta 3.4: Reducir las muertes prematuras por enfermedades no transmisibles
Base de datos:
WPRIM (Pacífico Occidental)
Asunto principal:
Arritmias Cardíacas
/
Alarmas Clínicas
/
Fatiga de Alerta del Personal de Salud
/
Unidades de Cuidados Intensivos
/
Monitoreo Fisiológico
Tipo de estudio:
Ensayo clínico controlado
Límite:
Humanos
Idioma:
Chino
Revista:
Chinese Journal of Medical Instrumentation
Año:
2020
Tipo del documento:
Artículo